EASYPULSE TOC BATTERY
ENSURE BATTERY IS FULLY CHARGED PRIOR TO FIRST USE.
NOTE: See initial charging instructions on back of Battery.
(DO NOT use the DC Power Cord for initial charge.)
Lithium Ion Battery Guidelines
Proper Use:
In order to prolong life of the battery,
it is best to charge the battery
frequently rather than waiting until
the battery is fully discharged.
Storage:
When storing for long periods of time
without use (> 1 month) be sure to
charge the battery to around 50%
charge level to prolong life of battery.
Battery
Handle
Battery ID
Charge Indicator
The Battery has it’s own Charge Indicator. Momentarily activating the PUSH
to Test Button will indicate the current state of battery charge. (Battery is
NOT shipped fully charged.)
Charge
Indicator
Push to Test
Button
4 LED Lights - 75% to 100% Full Charge
3 LED Lights - 50% to 75% Charge
2 LED Lights - 25% to 50% Charge
1 LED Light - 15% to 25% Charge
(Blinking < 15%)
Control Panel Battery Indicator
When the TOC is not connected to any external AC or DC power source it
will operate totally from the lithium-ion Battery.
It is imperative to monitor the status of the battery charge to ensure adequate
available battery power/runtime. (see “EasyPulse TOC User Interface” for
location of the Control Panel Battery Indicator).
4 Green LED’s 75% to 100% Full
3 Green LED’s 50% to 75% Full
2 Green LED’s 25% to 50% Full
1 Green LED
15% to 25% Full
1 Red LED
Less than 15% - Low Priority Alarm
Less than 7% - Medium Priority
Less than 2% - High Priority
LOW
NOTE:
Battery duration times will degrade with battery age, environmental
operating conditions, and use over time. If the battery is near the end
of its useful life, you may experience battery durations less than that