PRECISION DIGITAL PD233 Скачать руководство пользователя страница 2

– 15 –

Config Mode

Allows Scrolling Through CONFIG Menus

Engineering Units

Psi, ˝Hg, ˝H

2

O (with three temperature options, 60°F, 4°C

and 20°C), ft.SW, bar, mbar, kPa, MPa, mmHg, cmH

2

O,

mmH

2

O, kg/cm

2

Update Rate

4 Options: 10x/sec, 5x/sec, 2x/sec, 1x/sec

Auto Off

6 Options: Never, 2 Min., 5 Min., 15 Min., 30 Min., 

Dampening 6 Options: None, Average 2, 4, 6, 8 x update rate
Languages 7 Languages: English, Spanish, French, Italian, German,

Portuguese, Dutch

Backlite

5 Options: On/Off, 10 Sec., 30 Sec., 5 Min., 15 Min.

Field Recalibration

Zero & Span (password protected)

Contrast

7 Available Options

Disable Config Options

Allows disabling of Config Options (password protected)

Calibration Chart

10 Point Individual NIST Traceable Calibration Chart,

Standard

Standard Accessories

300 Series SS Protective Cover

Nylon Protective Carrying Pouch

Display

Description

Flashing

Gauge over/underpressured

0% or 100%

beyond 105% of range

LOW BAT

Low battery, replace

ERROR

Internal error, call customer

service (203) 378-8281

Pressure unit conversion exceeds

RES ERROR

display resolution or gauge

pressured beyond resolution

SPECIFICATIONS

WARNING AND ERROR MESSAGES

– 2 –

Congratulations on your purchase of the

Precision Digital Test Gauge with total error band
full-scale accuracy and the largest display readout
in the industry of .66˝ high. Other industry-lead-
ing features include twelve selectable engineering
units, seven languages, and password-protected
disable and calibration functions. With the range
printed on the keypad, this digital gauge meets
the ASME B40.7 digital gauge specification. See a
complete listing of product features and specifi-
cations on pages 14 & 15.

Precision Digital Corporation
19 Strathmore Road
Natick, MA 01760-2418
Tel: 800-343-1001
Fax: 508-655-8990

Содержание PD233

Страница 1: ...elow Pipe to which gauge is attached must be properly grounded 4 00 1 860 17 Battery Installation and Replacement The gauge comes standard with a quantity of three AAA alkaline batteries installed Use either Duracell MN2400 MX2400 or Energizer E92BP X92RP AAA alkaline non rechargeable batteries Batteries have a life of approximately 1000 hours Battery life is dependent on gauge usage backlite sett...

Страница 2: ...er Nylon Protective Carrying Pouch Display Description Flashing Gauge over underpressured 0 or 100 beyond 105 of range LOW BAT Low battery replace ERROR Internal error call customer service 203 378 8281 Pressure unit conversion exceeds RES ERROR display resolution or gauge pressured beyond resolution SPECIFICATIONS WARNING AND ERROR MESSAGES 2 Congratulations on your purchase of the Precision Digi...

Страница 3: ...F Dial Size 3 Case Material 300 Series SS Case Finish Electropolished Case Enclosure Rating Weatherproof IP65 Socket Material 316 SS Socket Size 1 4 NPT Male others on application Socket Location Lower connection 6 00 o clock Ranges Vac thru 7000 psi see other engineering units below for other units of measurement Operating Temperature 0 150 F Storage Temperature 40 180 F DISPLAY Type LCD Display ...

Страница 4: ... 000 4 Bargraph of full scale Flashing display when unit pressured below zero Press to indicate minimum or maximum pressure gauge has measured Press again to return to pressure units While in max or min mode press to clear minimum or maximum displayed values UNITS Pressure ENGLISH PSI INHG INH2O with Temperature Options 60 F 4 C 20 C FTSW METRIC BAR MBAR KPA MPA mmHG CMH2O mmH2O KG CM2 UPDATE RATE...

Страница 5: ...elect the desired option Step 5 Press ENTER to finalize the selection Notes Selecting the DISAB feature does not disable the CLEAR button on the keypad for the MAX MIN fea ture If the DISAB feature is selected pressing the ZERO button on the keypad will cause the display to read DISAB for two seconds The gauge will then revert back to the unit of measure of the gauge The DISAB feature disables the...

Страница 6: ... default and means the unit can be rezeroed between 5 of span If DISAB is selected the zero button is deactivated and no display change will occur when the zero button is pushed Step 1 Press the CONFIG key until the word ZERO SP appears Step 2 Press ENTER 6 Turns the gauge on and off When pressing the ON OFF key while in the off position gauge start up display first indicates the software version ...

Страница 7: ...e the gauge will turn itself off after no activity Indicates default 10 Step 1 Press the CONFIG key until the word CAL IBRAT appears Step 2 Press the ENTER key Step 3 The letters asterisks PW appear Step 4 Press the CONFIG key An Ø appears in the first position Step 5 Press the ENTER key once Step 6 Press the CONFIG key until Ø appears Ø will appear in the second position Step 7 Press ENTER Step 8...

Страница 8: ...lt password that allows access to gauge calibration To access the factory default password KEYPAD FUNCTIONS Indicates default Indicates default 8 Offerings are Never 2 5 15 or 30 minutes To use the AUTO OFF option Step 1 Press the CONFIG key until the word AUTO OFF appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select the desired AUTO OFF rate Step 4 Press the ENTER key to final...