Helios Large Display Temperature Meter Instruction Manual
PD2-7000
40
Advanced Features Menu
To simplify the setup process, functions not needed for most applications are located in the
Advanced
Features
menu.
Press and hold the Menu button for three seconds to access the advanced features of the meter.
Advanced Features Menu & Display Messages
The following table shows the functions and messages of the
Advanced Features
menu in the order they
appear in the menu.
Display
Parameter
Action/Setting
Adjust
Adjust
Set adjust value to offset temperature reading
T cal
T Cal
Enter temperature calibration menu
T lo
T Low
Apply the low temperature input (e.g. 32
°
F)
T Hi
T High
Apply the high temperature input (e.g. 1000
°
F)
rtdtot
RTD Total
Set the number of RTDs connected in parallel to obtain average
temperature
filtEr
Filter
Set noise filter value
bypASs
Bypass
Set filter bypass value
Round
Round
Set the rounding value for display variables
seriAl
Serial
Set serial communication parameters
SlaveId
Slave ID
Set slave ID or meter address
baud
Baud rate
Select baud rate
Tr dLY
Transmit delay
Set transmit delay for serial communication
Parity
Parity
Select parity: Even, Odd, or None with 1 or 2 stop bits
t-byt
Time byte
Set byte-to-byte timeout
SelEct
Select
Enter the Select menu (function, cutoff, out)