Model PD6060 Dual-Input Process Meter
Instruction Manual
84
Meter Copy or Cloning Instructions
Caution!
Do not connect the two meters to
the same signal source while
cloning. Internal calibration may
be affected.
1. Connect two meters using a PDA1200 meter copy cable.
Warning!
Using standard CAT5 or other cable will cause
damage to both meters. Use PDA1200 meter copy
cable only.
2. Do not connect the two meters to the same signal source.
3. Power up both meters. Leave Clone meter in Run Mode.
4. Enter
the
Advanced
Features
menu of the Master meter; see
Advanced Features Menu
on page 66.
5. Scroll to the
Copy
function using the Up
arrow button then press
Enter.
6. The meter displays the message
Send
. Press Enter, the display
flashes while sending data. The message
done
is displayed when
copying is completed.
7. The Clone meter displays the message
CPY Rc
while being
programmed then the message
done
when copying is completed.
The meter initializes and returns to Run Mode using the same
settings as the Master.
8. If meter to be cloned does not respond to the data being sent, refer
to Copy Function Requirements above.
!