20
21
Cold junction compensation ....................... < ±1.0°C
Sensor error detection ................................ Yes
Sensor error current:
When detecting .................................... Nom. 33 mA
Else
......................................................
0
mA
Voltage input:
Measurement range .................................... -12...150 mV
Min. span ..................................................... 5 mV
Input resistance ........................................... 10 M
Ω
Output:
Current output:
Signal range ................................................ 4...20 mA
Min. signal range ......................................... 16 mA
Updating time .............................................. 440 ms
Output signal at EEprom error ....................
≤
3.5 mA
Load resistance ...........................................
≤
(V
supply
- 7.2) / 0.023
[Ω]
Load stability .................................................... < ±0.01% af span / 100
Ω
Sensor error detection:
Programmable ............................................. 3.5...23 mA
NAMUR NE43 Upscale ............................... 23 mA
NAMUR NE43 Downscale ........................... 3.5 mA
Ex / I.S. data:
U i ................................................................. : 28 VDC
I i .................................................................. : 120 mADC
Pi ................................................................. :
0.84 W
L i ................................................................. : 10 µH
Ci ................................................................. : 1.0 nF
EEx / I.S. approval:
DEMKO 99ATEX126963 .............................. II 1 G
EEx ia IIC T1...T6
Max. amb. temperature for T1...T4 ............. 85°C
Max. amb. temperature for T5 and T6 ....... 60°C
Applicable in zone ....................................... 0, 1 or 2
Marine approval:
Det Norske Veritas, Ships & Offshore ......... Standard for Certification No. 2.4
GOST R approval:
Standard: VNIIM .......................................... Cert. no. Ross DK.ME48.V01899
Ex: VNIIFTRI ................................................ Cert. no. Ross DK.GB06.V00100
Observed authority requirements:
Standard:
EMC 2004/108/EC
Emission and immunity ....................... EN 61326
ATEX 94/9/EC .............................................. EN 50014 and EN 50020
Of span =