background image

3

OPERATION 

The 8072 is designed to sample process air at a pressure between -1 psig and +7psig. 
This range of air pressure is typical of that found in a plastics drying system.  If you are 
sampling compressed air, consult factory for other model options.  The plastic tubing 
supplied with this monitor can withstand temperatures up to 275ºF.  Higher temperature 
air can be cooled to a safe level by sampling through a short length of copper tubing. 

Older models with built-in copper cooling coils are supplied with clear PVC tubing.  
This tubing should be limited to 180ºF. 

NOTE:  For best sensor life and response time, avoid sampling ambient air of high 
humidity.  Leave unit off when not monitoring a dryer or hopper. 

Some units have a brass selector valve on the side panel.  This valve should remain 
closed during any operation to prevent ambient air from leaking into the sensor manifold. 

The 8072 is packaged for convenient spot checking of dryers or hopper systems, but it 
can be used continuously to monitor process conditions, with the beeper indicating 
excessive moisture. 

Dryer Output Sampling 

The performance of a dryer can be evaluated by measuring the dew point of its output 
air.   

1.

Connect the 8072 sample tubing to a point in the drier’s output.  Make sure not to
exceed the supplied tubing temperature rating.

2.

Turn on the 8072 and allow the reading to stabilize.  It can take 10 minutes or
more for the unit to stabilize when first used.  You can silence the beeper with the
switch on the side of the panel.

A constant high reading on the 8072 may indicate a malfunctioning dryer or one that is 
overloaded by damp material in the hopper. 

Hopper Output Sampling 

Dryness of material in a hopper can be checked by measuring the dew point of air exiting 
the hopper.  When a hopper is initially loaded with material, high dew points at the 
hopper air return are normal while moisture is removed.  It can take significant time to 
dry material in some systems. 

1.

Connect the 8072 sample tubing to a point in the hopper’s air output.  Make sure
not to exceed the supplied tubing temperature rating.

2.

Turn on the 8072 and allow the reading to stabilize.  Plastics processing can
proceed when the dew point reading is low enough for your material.

Отзывы: