14
Before doing any precision levelling it is advised to check the calibration of the unit.
‣
Set up the laser on a tripod at about 30m facing a wall or staff with a detector.
‣
Allow the unit to level.
‣
Detect and note laser position on the wall or staff.
‣
Without moving the tripod rotate the unit 180°.
‣
Detect and note laser position on the wall or staff.
‣
Calculate the distance between the two readings.
‣
The difference should be within 6mm at 30m.
‣
If the unit is out of calibration it is advised to send it in to Spot-on.
(see www.spoton.com.au for details)
CHECKING CALIBRATION