background image

Poseidon SE7EN User Manual

Appendix 3 Page 97

Figure 2-8.

 Tests 44 and 45, confirming sufficient gas supply pressures.

DANGER:

 

Always open the oxygen cylinder valve slowly. Rapid pressurization 
increases risk of fire. Opening the valve slowly reduces this risk. 
Carefully maintaining clean oxygen regulators, cylinders, and valves 
before, during, and after your dives will further reduce this risk. 

WARNING:

 

Once both cylinders have been turned on during Tests 44 and 45, 
do NOT turn them off again until after completing the dive. If they 
are turned off before completing the pre-dive routine, then Tests 
49 and/or 53 will fail. If they are turned off before the dive, the dive 
will be cut short. This is particularly true for the diluent cylinder, 
which provides breathing gas in the event of an emergency open-
circuit bailout.

Oxygen and diluent cylinder supplies (Test 44 & 45)

Tests 44 and 45 determine whether the Oxygen and Diluent cylinders, respectively, are turned 
on and have sufficient gas to conduct a dive. Following proper procedure, both cylinders will have 
been in the off position when Test 44 is reached (if not, gas will be wasted during Tests 24–27, 
which verify that the four solenoid valves draw the correct amount of power when held open). 

Each of these two tests will allow up to two minutes to turn on each cylinder. The bottom one, 
two, or three segments of the respective cylinder pressure bar graphs will flash until sufficient 
pressure is detected (Figure 2-8). When the system detects sufficient oxygen pressure, it then 
waits until it detects sufficient diluent pressure. Provided the oxygen cylinder pressure is 
greater than 34 bar / 493 psi, and the diluent pressure is greater than 51 bar / 739 psi, the 
automated pre-dive check will pass and the pre-dive test routine will continue. There is no 
upper limit for cylinder pressures for these two tests. However, it should be noted that the 
high-pressure sensors themselves have an upper limit to the pressure they can correctly read. 
The high-pressure sensor for the oxygen cylinder is limited to 207 bar / 3097 psi, and the 
sensor for the diluent is limited to 300 bar / 4410 psi. Exposing either sensor to a pressure in 
excess of these limits may yield unpredictable results. Also, oxygen pressures in excess of 
about 135 bar / 2000 psi pose a substantially increased risk of fire.

Содержание SE7EN

Страница 1: ...SE7EN connected to M28 POSEIDON SE7EN Appendix 3 version 1 2...

Страница 2: ...m allowed working pressure it holds 612 liter 21 6 cubic feet of air Because this cylinder is your open circuit OC bailout gas in the event of an emergency Poseidon strongly recommends that this cylin...

Страница 3: ...loop Repeat this procedure several times until you have pulled as strong a vacuum on the breathing loop as you can and then quickly switch the mouthpiece to OC position to hold the vacuum inside the...

Страница 4: ...on M28 then the first screen displayed shows the SE7EN mode display Figure 2 2 for a few seconds before switching automatically to initiate the PreDiveTest mode Figure 2 3 The firmware version number...

Страница 5: ...PSTs is as follows Test 1 Confirms the main data logger is functional Tests 2 9 verify the ROM RAM and EEPROM function in all four processors Test 10 verify the Bluetooth functionality Test 14 Confir...

Страница 6: ...The system allows the diver up to two minutes to make this switch The Closed Circuit icon will continue to display until Test 43 when the mouthpiece needs to be switched to the open circuit OC positio...

Страница 7: ...two decompression states do not exactly match exactly test 40 will fail Failure of this test is a notification to the diver that the system has detected this discrepancy between the two sets of decom...

Страница 8: ...the off position when Test 44 is reached if not gas will be wasted during Tests 24 27 which verify that the four solenoid valves draw the correct amount of power when held open Each of these two test...

Страница 9: ...racking pressure of the over pressure relief valve on the bottom of the right counterlung when that valve is adjusted to its maximum cracking pressure The system then monitors the loop pressure for th...

Страница 10: ...mouthpiece is in the Closed Circuit position and is detected by the HUD the test will pass The system allows 2 minutes for this test to be completed before timing out As with the Open Circuit mouthpie...

Страница 11: ...the RBConfig PC software Prebreathe Figure 2 13 Prebreathe Open circuit regulator function Test 54 After completing Test 53 the information field will have the text Switch mouthpiece to Open Circuit...

Страница 12: ...mouthpiece tighten Attach HUD on to mouthpiece Attach LP diluent supply hose to inflator Close OPV on right counterlung Negative loop pressure test Insert smart battery and conduct power up self test...

Страница 13: ...position NOW In most cases this signal will be triggered in association with an open circuit bailout situation thereby instructing the diver to switch the mouthpiece from closed circuit mode to open c...

Страница 14: ...to alert other nearby divers of a potential problem Monitoring the M28 primary display Most of the information concerning the status of the dive and the various system parameters is communicated to t...

Страница 15: ...etail Figure 3 2 LCD display layout Figure 3 3 Symbols DANGER If the Primary Display screen is ever blank while diving the Poseidon SE7EN immediately commence an abort to the surface in Open Circuit m...

Страница 16: ...r of the screen means all systems are functioning properly and all parameters are working cor rectly In most cases the signals will flash when activated further drawing attention Abort and open circui...

Страница 17: ...ive should be termi nated whenever this icon is displayed The diver should ascend towards the surface at a slow and controlled rate watching the Primary Display for the Stop Alert and additional decom...

Страница 18: ...Chapter 2 the linearity of the oxygen sensor response is only validated up to a PO2 value of 1 0 bar atm i e 100 oxygen at sea level Most rebreathers assume that the sensor response remains linear at...

Страница 19: ...C0 C1 C2 Mouthpiece position The difference between NC no circuit and UN unknown depends on whether the mouthpiece is reporting that neither closed circuit nor open circuit is currently established no...

Страница 20: ...21 Current depth Maximum depth Ceiling Figure 3 22 Maximum depth The maximum depth achieved during the dive is displayed above the current depth in the lower left part of the Primary Display Ceiling i...

Страница 21: ...26 Elapsed dive time The number of minutes that have elapsed during the dive i e the total dive time is dis played in the lower right part of the Primary Display above the RDT value This value repre...

Страница 22: ...icator If the battery fails the entire life support system including alarms may cease to function Failure to abort to open circuit and terminate the dive could lead to serious injury or death Cylinder...

Страница 23: ...ltering the center icon see figure 3 35 This selection will remain as long as you stay in the OC mode If you decide to go back to CC the gas will automatically be restored to your SE7EN predefined dil...

Страница 24: ...t the Air Dil pressure graph is not flashing The electronics system will constantly calculate whether there is enough air supply remaining to allow a safe open circuit bailout to the surface If there...

Страница 25: ...tments Besides the three large straps for attachment to the harness each counterlung has several additional straps used to adjust positioning At the top of each counterlung is a single adjust able str...

Страница 26: ...specially deep inhaled breath For minor adjustments in buoyancy it is usually easier to control it with your breath rather than the purge button since it could easally give a bit too much gas To decre...

Страница 27: ...following four conditions have all been met the depth is 0 the back of the Primary Display where the wet switch contacts are located has been dried the pressure in the diluent regulator and hoses has...

Страница 28: ...l continue to flash and there may be a symbol in the upper right field of the display flashing to indicate the nature of the problem For example the directional arrows advise you to go up ascend if yo...

Страница 29: ...the physical product needs to be connected and then the Power out option needs to be enabled To enable the power out function goto the SYSTEM app and press both buttons Select Connect and then Power...

Страница 30: ...1 6 l CO2 per minute at STPD Standard Temperature Pressure Dry The SE7EN can support decompression diving To be able to perform decompression dives you need two things 1 the appropriate training and 2...

Страница 31: ...of the counterlungs to the gas processor Figure 1 16 Three velcro straps on the back of each counterlung Figure 1 16 attach the counterlungs to the shoulder straps of the backpack The SE7EN is provid...

Страница 32: ...s These valves can be replaced with blank covers if manual gas addition capabilities are not required e g for recreational diving The port for the manual Oxygen addition valve right counterlung is int...

Страница 33: ...tor hose to your onboard diluent gas supply or any other offboard source Route the hose over your left sholder and connet to the diluent inflator valve though the quick connection The valves described...

Страница 34: ...has been reached The generated alarm response will be similar to the alarm shown when the diver exceeds the maximum working depth of the rig This value also affects the way the controlling resource al...

Страница 35: ...the diluent composition to air If the Recreational 40m battery or Deco 40m battery was inserted by mistake the diver should NOT confirm these tests Instead the test should be allowed to time out two m...

Страница 36: ...ill not issue an alarm HUD Buddy lights Audio Alarm when decompression is entered There is no audio alarm when decompression is entered on a non decompres sion configuration When the mouthpiece is fou...

Отзывы: