background image

Poseidon SE7EN User Manual

Appendix 3 Page 100

Figure 2-12.

  

Test 55 – Service interval check.

Ready to dive

Under normal circumstances the Poseidon SE7EN will successfully complete all Pre-Dive 
checks in about three minutes, and a screen will appear on the display that looks like that 
shown in Figure 2-14, with a PO

2

 value between 0.3 and 0.9 (usually the number is towards 

the higher end of this range because the loop is partially filled with oxygen during Tests 49 and 
53), a depth of 0, a dive time of 0, and a remaining dive time clock showing 199 minutes.

Figure 2-14.

 Ready-to-dive.

At this point,the pre-dive tests are 
completed, and the system is ready to dive.

Service interval check (Test 55)

The final test (Test 55; Figure 2-12) is also the simplest. This test merely ensures that the 
rebreather unit does not require servicing. Each rebreather unit must be brought to a qualified 
Poseidon Service Center at least once every two years, to receive updates and make any 
necessary repairs or adjustments. When Test 55 is displayed, the information field will say 
“Service in X weeks” which is the number of weeks remaining before servicing will be required. 
When this value gets low, return the rebreather to an authorized Poseidon service center for 
maintenance. Upon passing Test 55, the pre-dive routine is complete. 

Once the service date is due, the diver is prompted to acknowledge with YES or NO that he/she 
has understood that service is needed. A grace period of 4 weeks is added.
The service due date can also be seen through the RBConfig PC software.

Prebreathe

Figure 2-13.

 Prebreathe.

Open circuit regulator function (Test 54)

After completing Test 53, the information field will have the text “Switch mouthpiece to Open 
Circuit” indicating the need change the mouthpiece back to OC mode. When the mouthpiece 
is switched, the sentence Breath through the mouthpeice are shown on the screen. This 
prompts the diver to test the function of the open-circuit regulator. After inhaling several 
breaths from the regulator, the test passes.

Directly after Pre-Dive has completed, the diver will be prompted to perform a prebreathe. This 
is shown in the information box in the main screen. When the user switches the mouthpiece to 
CC mode, the remaining prebreathe counter will count down from 300 seconds. If the user 
switches out of CC then the counter will pause. Once the counter has counted down to zero, 
the rig enters ready to dive mode. It will stay in this state for 30 minutes. If the dive is not 
started in this time, then it goes back to prebreathe mode. Prebreathe mode is not mandatory 
and can be terminated without penalty by starting the dive. 30 minutes after the end of a dive , 
the rig will once again enter prebreathe mode. Note that during prebreathe mode, the rig will fire 

the diluent and oxygen solenoids for 2 seconds close to the start of each prebreathe session, if 
the diver is not breathing from the loop at this stage the unit may issue an alarm. 

The reason for using the solenoids at this stage is that if a diver has turned off either of the cyl-
inders, this will result in a large pressure drop which will prompt the diver to turn the cylinder on. 
If a closed cylinder is detected during prebreathe then the appropriate bar graph will flash.

Содержание SE7EN

Страница 1: ...SE7EN connected to M28 POSEIDON SE7EN Appendix 3 version 1 2...

Страница 2: ...m allowed working pressure it holds 612 liter 21 6 cubic feet of air Because this cylinder is your open circuit OC bailout gas in the event of an emergency Poseidon strongly recommends that this cylin...

Страница 3: ...loop Repeat this procedure several times until you have pulled as strong a vacuum on the breathing loop as you can and then quickly switch the mouthpiece to OC position to hold the vacuum inside the...

Страница 4: ...on M28 then the first screen displayed shows the SE7EN mode display Figure 2 2 for a few seconds before switching automatically to initiate the PreDiveTest mode Figure 2 3 The firmware version number...

Страница 5: ...PSTs is as follows Test 1 Confirms the main data logger is functional Tests 2 9 verify the ROM RAM and EEPROM function in all four processors Test 10 verify the Bluetooth functionality Test 14 Confir...

Страница 6: ...The system allows the diver up to two minutes to make this switch The Closed Circuit icon will continue to display until Test 43 when the mouthpiece needs to be switched to the open circuit OC positio...

Страница 7: ...two decompression states do not exactly match exactly test 40 will fail Failure of this test is a notification to the diver that the system has detected this discrepancy between the two sets of decom...

Страница 8: ...the off position when Test 44 is reached if not gas will be wasted during Tests 24 27 which verify that the four solenoid valves draw the correct amount of power when held open Each of these two test...

Страница 9: ...racking pressure of the over pressure relief valve on the bottom of the right counterlung when that valve is adjusted to its maximum cracking pressure The system then monitors the loop pressure for th...

Страница 10: ...mouthpiece is in the Closed Circuit position and is detected by the HUD the test will pass The system allows 2 minutes for this test to be completed before timing out As with the Open Circuit mouthpie...

Страница 11: ...the RBConfig PC software Prebreathe Figure 2 13 Prebreathe Open circuit regulator function Test 54 After completing Test 53 the information field will have the text Switch mouthpiece to Open Circuit...

Страница 12: ...mouthpiece tighten Attach HUD on to mouthpiece Attach LP diluent supply hose to inflator Close OPV on right counterlung Negative loop pressure test Insert smart battery and conduct power up self test...

Страница 13: ...position NOW In most cases this signal will be triggered in association with an open circuit bailout situation thereby instructing the diver to switch the mouthpiece from closed circuit mode to open c...

Страница 14: ...to alert other nearby divers of a potential problem Monitoring the M28 primary display Most of the information concerning the status of the dive and the various system parameters is communicated to t...

Страница 15: ...etail Figure 3 2 LCD display layout Figure 3 3 Symbols DANGER If the Primary Display screen is ever blank while diving the Poseidon SE7EN immediately commence an abort to the surface in Open Circuit m...

Страница 16: ...r of the screen means all systems are functioning properly and all parameters are working cor rectly In most cases the signals will flash when activated further drawing attention Abort and open circui...

Страница 17: ...ive should be termi nated whenever this icon is displayed The diver should ascend towards the surface at a slow and controlled rate watching the Primary Display for the Stop Alert and additional decom...

Страница 18: ...Chapter 2 the linearity of the oxygen sensor response is only validated up to a PO2 value of 1 0 bar atm i e 100 oxygen at sea level Most rebreathers assume that the sensor response remains linear at...

Страница 19: ...C0 C1 C2 Mouthpiece position The difference between NC no circuit and UN unknown depends on whether the mouthpiece is reporting that neither closed circuit nor open circuit is currently established no...

Страница 20: ...21 Current depth Maximum depth Ceiling Figure 3 22 Maximum depth The maximum depth achieved during the dive is displayed above the current depth in the lower left part of the Primary Display Ceiling i...

Страница 21: ...26 Elapsed dive time The number of minutes that have elapsed during the dive i e the total dive time is dis played in the lower right part of the Primary Display above the RDT value This value repre...

Страница 22: ...icator If the battery fails the entire life support system including alarms may cease to function Failure to abort to open circuit and terminate the dive could lead to serious injury or death Cylinder...

Страница 23: ...ltering the center icon see figure 3 35 This selection will remain as long as you stay in the OC mode If you decide to go back to CC the gas will automatically be restored to your SE7EN predefined dil...

Страница 24: ...t the Air Dil pressure graph is not flashing The electronics system will constantly calculate whether there is enough air supply remaining to allow a safe open circuit bailout to the surface If there...

Страница 25: ...tments Besides the three large straps for attachment to the harness each counterlung has several additional straps used to adjust positioning At the top of each counterlung is a single adjust able str...

Страница 26: ...specially deep inhaled breath For minor adjustments in buoyancy it is usually easier to control it with your breath rather than the purge button since it could easally give a bit too much gas To decre...

Страница 27: ...following four conditions have all been met the depth is 0 the back of the Primary Display where the wet switch contacts are located has been dried the pressure in the diluent regulator and hoses has...

Страница 28: ...l continue to flash and there may be a symbol in the upper right field of the display flashing to indicate the nature of the problem For example the directional arrows advise you to go up ascend if yo...

Страница 29: ...the physical product needs to be connected and then the Power out option needs to be enabled To enable the power out function goto the SYSTEM app and press both buttons Select Connect and then Power...

Страница 30: ...1 6 l CO2 per minute at STPD Standard Temperature Pressure Dry The SE7EN can support decompression diving To be able to perform decompression dives you need two things 1 the appropriate training and 2...

Страница 31: ...of the counterlungs to the gas processor Figure 1 16 Three velcro straps on the back of each counterlung Figure 1 16 attach the counterlungs to the shoulder straps of the backpack The SE7EN is provid...

Страница 32: ...s These valves can be replaced with blank covers if manual gas addition capabilities are not required e g for recreational diving The port for the manual Oxygen addition valve right counterlung is int...

Страница 33: ...tor hose to your onboard diluent gas supply or any other offboard source Route the hose over your left sholder and connet to the diluent inflator valve though the quick connection The valves described...

Страница 34: ...has been reached The generated alarm response will be similar to the alarm shown when the diver exceeds the maximum working depth of the rig This value also affects the way the controlling resource al...

Страница 35: ...the diluent composition to air If the Recreational 40m battery or Deco 40m battery was inserted by mistake the diver should NOT confirm these tests Instead the test should be allowed to time out two m...

Страница 36: ...ill not issue an alarm HUD Buddy lights Audio Alarm when decompression is entered There is no audio alarm when decompression is entered on a non decompres sion configuration When the mouthpiece is fou...

Отзывы: