![POSDION REXTAR X Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/posdion/rextar-x/rextar-x_operation-and-service-manual_1582103031.webp)
POSDION Co., Ltd.
REXTAR-X Manual
PN# 100304-00
Rev.02
Page
:
25
8. Operation
1) Prepare for Operation
(1) Lead aprons must be
worn by patients
during Exposure.
(2) In order to minimize the time required per session and gain the best results, the device must be
kept from being shaken or otherwise disturbed
when pressing the exposure button
.
(3) In order to set the wanted sec value free of error, pay attention to the decimal point when setting
the sec value
. Be sure that the leaded scatter shield is in place
.
(4) Visitors must be sent outside of the room while X-RAY filming patients.
(5) Close and continuous attention must be paid to the maintenance of the device.
(6) Accumulation of radiation should not exceed the maximum recommended level.
Especially if long time intervals are often used, it is necessary that an expert assess the situation
and determine if the user needs to take any extra protective measures.
※
WARNING
※
It is imperative to check that the frequency and voltage of power
feeding equipment follow specifications written on the system
labels affixed to the body of the device.
Voltage must be within ± 10% of the normal level.
Содержание REXTAR X
Страница 2: ...This page intentionally left blank...
Страница 4: ......
Страница 6: ...This page intentionally left blank...
Страница 30: ...POSDION Co Ltd Page 24 Rev 02 PN 100304 00 REXTAR X Manual 3 Tank Label Pic1 7 2 3 TANK Label bottom...
Страница 52: ...POSDION Co Ltd Page 40 Rev 02 PN 100304 00 REXTAR X Manual 3 Block diagram...
Страница 55: ...POSDION Co Ltd REXTAR X Manual PN 100304 00 Rev 02 Page 43 For Authorized Service and Test Technicians...
Страница 56: ...POSDION Co Ltd Page 44 Rev 02 PN 100304 00 REXTAR X Manual For Authorized Service and Test Technicians...
Страница 57: ...POSDION Co Ltd REXTAR X Manual PN 100304 00 Rev 02 Page 45 For Authorized Service and Test Technicians...
Страница 58: ...POSDION Co Ltd Page 46 Rev 02 PN 100304 00 REXTAR X Manual For Authorized Service and Test Technicians...
Страница 59: ...POSDION Co Ltd REXTAR X Manual PN 100304 00 Rev 02 Page 47 For Authorized Service and Test Technicians...
Страница 60: ...POSDION Co Ltd Page 48 Rev 02 PN 100304 00 REXTAR X Manual For Authorized Service and Test Technicians...
Страница 61: ...POSDION Co Ltd REXTAR X Manual PN 100304 00 Rev 02 Page 49 For Authorized Service and Test Technicians...
Страница 62: ...POSDION Co Ltd Page 50 Rev 02 PN 100304 00 REXTAR X Manual For Authorized Service and Test Technicians...
Страница 63: ...POSDION Co Ltd REXTAR X Manual PN 100304 00 Rev 02 Page 51 This page intentionally left blank...