Recalibrating Temperature Controller
Even though the temperature settings have been set under controlled laboratory conditions,
it is recognised that they may need to be adjusted to compensate for local environmental
conditions. By carrying out the following procedure
the operator may introduce an „offset‟
temperature adjustment to suit actual conditions.
Important Note:
The following procedure must be carried out by a suitably qualified
technician. Failure to follow these instructions exactly, or the
adjustment of other undocumented parameters, could render the
instrument inoperable resulting in the instrument being returned to the
manufacturer for repair and re-calibration.
Important Note:
Please consult your suppliers or Porvair Sciences directly if you
require to carry out recalibration of older TLK-38 or TLK-31
temperature controllers.
For example: If the temperature displayed is 60°C and the temperature measured from the needles is
55°C then an offset of „5
‟
will typically be required.
After adjustment the temperatures must be allowed to stabilise for a few minutes before checking. All
gas temperature measurements should be carried out using a suitable calibrated thermo-probe. For
further information, please contact your supplier or Porvair Sciences direct.
K32
Press and hold ‘
P
’ for 3 seconds.
Hold ‘
Up
’ Arrow until 1400 is reached.
Press ‘
P
’
Press ‘
U
’ eleven times until ‘
cAL
’ is displayed.