PW336i
Hardware User Manual
38
13
.
How to output a greater current than allowed max. single phase current?
The three-phase current power of kit can be used by parallel connection. Users need set the
phase output. It is suggested to set amplitude the same.
The total output amplitude of phase parallel connection is: I=Ia+Ib+Ic=30 + 30 + 30= 90A
14
.
How to output a greater voltage than allowed max. single phase voltage?
R
L
Load
Fig. 1-9
When the two voltages are parallel connected, the Ua and Ub amplitudes are the same and the
phases are opposite. The max AC voltage 240V can be got, and the max output current is still
0.5A-0.55A.