2.10 Environment Conditions and Factors
The basic working conditions are for factory inspection, and the rated working conditions are for
products adjustment and usage environment.
Factor
Basic working condition
Rated working condition
Environment
temperature
20
℃
±2
℃
-5
℃~
+45
℃
Relative humidity
45%
~
75%
10%
~
90%
Pressure
86kPa
~
106kPa
80kPa
~
110kPa
AC
power
supply
Voltage
220V±1%
220V±10%
Frequency
49.5Hz
~
50.5Hz
49Hz
~
51Hz
Waveform
Sine wave
2% of the harmonic distortion is
allowed.
Sine wave
The harmonic distortion is no
more than 5%.
It applies the temperature compensation technology. Under the rated working conditions, the
output accuracy will keep still between 15
℃~
35
℃
. The additional effects of temperature change
to output accuracy will be ±0.012%RD / K out of the range of rated
Under rated working condition, the frequency change of power supply will not affect technical
performance indications of current and voltage. The power factors between voltage change of
power supply to output power of AC current and voltage should meet following requirements.
Power
supply
Max. output power
Max. load voltage
Max. load current
AC current (6A)
AC voltage(120V)
DC current (20mA)
DC voltage (±10V)
198V
24VA
27.5VA
1)
10V
±500mA
220V
30VA
30VA
10V
±500mA
242V
30VA
30VA
10V
±500mA
Note 1: When the power supply is 198V and the max. AC voltage output is 110V, the max. output
power is 27.5VA