PM605A Hardware Instruction Manual 

 

 

 

 

 

 

 

 

VERSION: 

 

 

 

 

PM605A-AE-1.10

 

DATE: 

 

 

 

 

DEC. 2019 

 

 

This manual is the publisher of PONOVO POWER CO.,  LTD. To make any kind of copy of this 

manual please contact PONOVO POWER CO., LTD in advance. 

 

This manual represents the technical status for the moment of publishing. The product information, 

description and specifications mentioned in the manual do not have any contact binding force and 

PONOVO POWER CO., LTD remains the right to make modifications to the technical 

specifications and configurations without prior notice. PONOVO POWER CO., LTD does not take 

responsibility to the possible error/mistakes in this manual.

 

PONOVO POWER CO,. LTD

 

 

NO. 139 JINGHAI THIRD ROAD, BDA, BEIJING, CHINA, 100176 

 

Office 

TEL. +86 (10) 59089666 

 

E-Mail 

[email protected] 

 

Website 

www.relaytest.com / www.ponovo.com.cn 

 

PM605A 

HARDWARE INSTRUCTION MANUAL

 

Содержание PM605A

Страница 1: ...tion and specifications mentioned in the manual do not have any contact binding force and PONOVO POWER CO LTD remains the right to make modifications to the technical specifications and configurations without prior notice PONOVO POWER CO LTD does not take responsibility to the possible error mistakes in this manual PONOVO POWER CO LTD NO 139 JINGHAI THIRD ROAD BDA BEIJING CHINA 100176 Office TEL 8...

Страница 2: ...nt static electricity the tester must be connected to ground before testing 3 Any connector in front panel is not allowed to connect with ground 4 When the device is working the output signal cable is prohibited to plug 5 Disconnect the power supply before disconnecting cables 6 Avoid electric shock accident when the output voltage is above 36V 7 It s not allowed to feed external voltage current i...

Страница 3: ...7 2 9 FAULT PROTECTION 17 2 10 ENVIRONMENT CONDITIONS AND FACTORS 18 2 11 SHELL PROTECTION 19 2 12 COMMUNICATION INTERFACE 19 2 13 OTHER PARAMETERS 19 3 OFF LINE WORKING METHOD 20 3 1 GENERAL DESCRIPTION 20 3 2 OPERATION PANEL 20 3 3 OFF LINE INTERFACE 21 3 4 AC FIXED POINT CALIBRATION 22 3 5 DC FIXED POINT CALIBRATION 25 3 6 AC MANUAL CALIBRATION 26 3 7 BINARY CALIBRATION 27 3 8 AC U4 CALIBRATION...

Страница 4: ...PM605A Hardware Instruction Manual 4 APPENDIX C PGPS02 DIAGRAM 39 ...

Страница 5: ...c temperature compensation technology It makes that the setting value is standard value including phase and power and output accuracy keeps still with temperature 15 35 etc come true The PM605A software includes many test functions DC sampling auto calibration synchronization test transmitter auto calibration system function test harmonic test and manual test unit 1 2 System Configuration Host com...

Страница 6: ...CD screen using while offline 6 Function button using while offline 7 USB or Ethernet data interface 8 Rotary encoder setting parameters while offline 9 Signal indicator POWER POWER indicator lights when the power is supplied RUN RUN indicator flickers or off when the kit is working ALARM ALARM indicator flickers and gives off beep alarm signal when the device is over hot or overload during output...

Страница 7: ...inal A B C D E 2 Binary output terminal 1 2 3 4 3 Separated 300V DC voltage output terminal Optional The indicator will light when the output is beyond safe voltage 4 GPS synchronization signal input terminal 5 Power pulse input output terminal 6 Ventilation 7 ON OFF 8 Grounding terminal 9 Power ...

Страница 8: ...corresponding to the CH1 CH2 and CH3 of current on the panel The output amplitude and phase are independent Output range 0 6A Output resolution and min adjustment step 0 1mA Accuracy 0 05 The relative error is no more than 0 05 within 100mA 6A The absolute error is no more than 50μA within 0 100mA Stability 0 01 3min Max power output 30VA phase Harmonic distortion THD 0 05 ...

Страница 9: ...on the panel The output amplitude and phase are independent Output range 0 120V Output resolution and min adjustment step 1mV Accuracy 0 05 The relative error is no more than 0 05 within 5V 120V The absolute error is no more than 50μA within 0 100mA Stability 0 01 3min Max power output 30VA phase The output power is more than 15VA when the output is 57 7V Harmonic distortion THD 0 05 ...

Страница 10: ... 1mV Accuracy 0 1 The relative error is no more than 0 1 within 5V 120V The absolute error is no more than 10mV within 0 5V Stability 0 05 3min Max power output 30VA Harmonic distortion THD 0 1 2 1 4 AC Frequency and Phase AC frequency Output range 45Hz 65Hz Output resolution and min step 0 001 Hz Accuracy 0 001Hz Stability 0 0005 Hz 3min AC phase Output range 0 360 Output resolution and min step ...

Страница 11: ...tor adjustment range 1L 0 1C Accuracy 0 0005 Stability 0 0001 3min Active power and inactive power Accuracy 0 05 Stability 0 01 3min Active energy and inactive energy Level 0 05 Harmonic Harmonic number 2 21 Harmonic content 0 20 Harmonic phase 0 360 Amplitude accuracy 0 2 Phase accuracy 0 2 ...

Страница 12: ...solute error is no more than 0 5mA within 0 5mA Stability 0 05 3min Max output power 30W Max terminal voltage 6V Ripple content 0 2 DC output from AC voltage channel Output channel U1 of AC voltage channel as positive of DC voltage U2 as negative Output range 0 300V Output resolution and min adjustment step 1mV Accuracy 0 1 The relative error is no more than 0 1 within30V 300V The absolute error i...

Страница 13: ...The absolute error is no more than 0 1μA within 0 0 2mA Stability 0 01 3min Max Output power 0 2W Max load voltage 10V Ripple content 0 1 DC voltage Output range 0 10V Output resolution and min adjustment step 0 1mV Accuracy 0 05 The relative error is no more than 0 05 within 50mV 10V The absolute error is no more than 25μV within 0 50mV Stability 0 01 3min Max output power 5W Max load current 500...

Страница 14: ...C current measurement Input range 20mA Out of range 120 Resolution 0 001 Display digits 5 2 1 or 6 2 1 20mA range Accuracy 0 01 RG DC voltage measurement Input range 10V Out of range 120 Resolution 0 001 Display digits 5 2 1 or 6 2 1 10V range Accuracy 0 01 RG ...

Страница 15: ...ess than AC DC 250V 500mA The time error of binary output 0 1ms Editable pulse output frequency 0 500Hz adjustable pulse width 1ms 500ms adjustable 2 5 2 Binary Input 5 pairs independent binary inputs The separated voltage is 500V Empty contact and active contact 30V 250V inputs automatically recognizing the input polarity of active contact Max input voltage 250V DC Binary action time error 1ms ...

Страница 16: ...me Synchronization Input 2 7 1 PGPS Synchronization Device Signal Input Signal interface of PGPS synchronization instrument Input of pulse per minute and pulse per second signal type TTL TTL signal level 5V 2 7 2 Synchronization Time Accuracy PPM pulse per minute and PPS pulse per second are the time standard The pulse edge and input signal source are synchronized The time error 10ns Time error of...

Страница 17: ... current 100mA Ripple content 0 2 2 9 Fault Protection Current output overload protection function when the voltage output is short circuit or overloaded the kit will make alarm and the light flash Over hot protection function when it is over hot the kit will make alarm and the light will flash After alarm it will stop output automatically Auto current open circuit protection and current open circ...

Страница 18: ...tion technology Under the rated working conditions the output accuracy will keep still between 15 35 The additional effects of temperature change to output accuracy will be 0 012 RD K out of the range of rated Under rated working condition the frequency change of power supply will not affect technical performance indications of current and voltage The power factors between voltage change of power ...

Страница 19: ...ard applies with the IP21 regulated in GB 4208 1993 2 12 Communication Interface RS 232 interface synchronized with PGPS PC communication interface USB interface or RJ45 interface 2 13 Other Parameters Aluminum case with cover and handle Dimension 360mm 160mm 367mm W H D Weight 15Kg ...

Страница 20: ...M605A operation panel The operation panel includes LCD monitor rotary encoder and five function buttons Main off line operation Parameters are set by rotary encoder The items can be switched by rotary encoder Under a certain item press the encoder then rotate it It can increase or decrease the values or switch into other items Integrated with the soft keyboard some special setting functions can al...

Страница 21: ...tomatically enters into the following interface after power on Figure 3 2 Main interface of off line Operate the rotary encoder move the cursor into the item needed and press the encoder enter into the selected interface In any unit we can come back to the main interface by pressing exit button ...

Страница 22: ...rface Calibration interface includes Right of the interface Users can set calibration parameters here Left of the interface it includes AC voltage current amplitude phase AC voltage three phase active power reactive power and apparent power Users can press the third button to display AC output UI or AC power PQ and press the fourth button to display power of three phase four line 3P4L or three pha...

Страница 23: ...ion power 3 Frequency calibration Set AC frequency Put out calibration power 4 Power factor calibration Set power factor active power factor cosφ calibration points including 0 000L 0 500L 0 866L 1 000 0 866C 0 500C 0 000C reactive power factor sinφ calibration points including 1 000L 0 866L 0 500L 0 000 0 500C 0 866C 1 000C Put out calibration power 5 AC power calibration Set AC current output Se...

Страница 24: ...e DC amplitude will be showed in DC ln area and percentage error will be showed in Err area 2 Voltage transmitter Set AC voltage output Select voltage in transmission type Select signal output range of transmitter 0 10V 0 5V 0 1V Select signal output method of transmitter unidirectional bidirectional symmetrical Connect DC signal output of transmitter with PM605A DC signal input The DC amplitude w...

Страница 25: ...urrent values listed based on user setting Left corner of the interface The DC measurement result is showed in DC ln area Users press the second button to select the DC current measurement value put in IdcL or DC voltage measurement value put in Udc The calibration error is showed in Err area 3 5 1 DC Voltage and Current Calibration Set DC voltage 10 10V DC current 0 20mA Put out calibration power...

Страница 26: ...1 000 cosΦ2 1 000 cosΦ3 0 000 P1 W 0 000 P2 W 0 000 P3 W 0 000 Q1 Var 0 000 Q2 Var 0 000 Q3 Var Figure 3 5 AC manual calibration interface Calibration interface includes Left users can set calibration parameters Right every phase power factor active power reactive power listed according to users setting 3 6 1 AC Voltage Current and Phase Calibration Users can set any AC output current and voltage ...

Страница 27: ...rs can set calibration parameters Right it is binary input status and action time measurement result 3 7 1 Remote Signal Function Test Set binary output reverse status and time Connect binary output signal with measurement device input the binary signal need extra voltage Output check remote signal drift 3 7 2 Incident Resolution Test SOE Set 2 channels above binary outputs pre reverse status and ...

Страница 28: ...ameters of 4 channel binary output reverse status and time the same Extend 4 channels binary output signals into multi channel parallel outputs and connect with remote signal input terminal The binary output needs extra voltage Output check the capability of solve event avalanche ...

Страница 29: ... U3 V 57 735 57 735 57 735 240 00 0 00 120 00 U4 V 57 735 0 00 Φ1 0 00 Φ2 0 00 Figure 3 7 AC Test 380V interface AC Test 380V interface includes Right users can set calibration parameters Left Four phase voltage amplitudes and phase output values are based on the users setting Lower left DC measurement result is showed in DC ln area User presses the second button to select DC current measurement v...

Страница 30: ...rr 0 0000 I1 A I2 A 0 0000 0 0000 I3 A 0 0000 Figure 3 8 DC Test 5A interface DC Test 5A interface Right users can set calibration parameters Left Four phase voltage amplitudes and phase output values are based on the users setting Lower left DC measurement result is showed in DC ln area User presses the second button to select DC current measurement value input IdcL or DC voltage measurement inpu...

Страница 31: ... mA Err 0 0000 U1 2 V U300 V 0 000 0 000 Figure 3 9 DC 300V calibration interface DC Test 300V interface includes Right users can set calibration parameters Left Four phase voltage amplitudes and phase output values are based on the users setting Lower left DC measurement result is showed in DC ln area User presses the second button to select DC current measurement value input IdcL or DC voltage m...

Страница 32: ...D is binary output the dotted line with arrowhead in the left will be connected with binary While calibrating power transmitter the dotted line with arrowhead in the right will be connected with DC 3 11 2 Calibration Operation 1 Entering into calibration function interface In the main interface select function unit by rotary encoder and enter into function unit interface 2 Setting parameters Selec...

Страница 33: ...tput After setting parameters press function key 1 the kit will start to work The indicator Run on the panel will flash according to set parameters and status When it displays stop on the function key 1 users press key 1 to stop working During output process the output parameters can also be changed via rotating the encoder ...

Страница 34: ...n might be The power supply is too low Measuring the power supply by multimeter if the voltage is less than 170V the power supply should be improved There is no indication after switching on the kit The reason might be Power insurance blown 220V power supply fails to connect If there is sound of the fan the power indicator might be broken ...

Страница 35: ...e Indicate the GPS Leap Seconds and present time validation Time zone setting change the local time zone to display the right time Display local geography position coordinates 2 Specifications Timing Pulse Signal Pulse Signal Level TTL or RS 232 Timing Error between Two PGPS02 GPS TYP 100ns MAX 500ns Pulse Polarity Positive Pulse Width 100ms GPS Locking Time Local First Power on 34s 50s Restart Af...

Страница 36: ...Input Voltage DC 6 V 12 V Input Current 500mA Power Supply by RT Device DC 5 V Environmental Operating Temperature 20 c 70 c Storage Temperature 30 c 80 c General Weight 640g PGPS02 GPS Only Dimensions W H D 95mm 45mm 160mm PGPS02 GPS Only ...

Страница 37: ...p scintillating per second after the device synchronized with GPS 3 MODE Key Pressed to shift operating interfaces 4 SYN Key Synchronous Lamp Pressed the SYN key the synchronous test starts and the device will output a sync pulse at the setting time then the test stopped the lamp lights up until the test stop During the test procedure press the SYN key again will stop the test 5 Up Key Right Key P...

Страница 38: ...power socket 9 Serial Port Connector to RT Test Instrument Type of port DB9 F J 1 Power supply 5V J 2 READY Synchronous pulse output TTL level J 3 PPS Synchronous second pulse output TTL level J 5 GND Ground J 9 GND Ground Others are NC 10 Serial Port Connector to Computer Type of port DB9 M J 2 READ Serial in J 3 WRITE Serial out J 5 GND Ground J 8 PPS Synchronous second pulse output RS232 lever ...

Страница 39: ...PM605A Hardware Instruction Manual 39 Appendix C PGPS02 Diagram GPS Antenna GPS Clock Serial Port Line Note DC power is supplied while PGPS02 fails to provide power ...