B-3
B Optional Accessories for the Sensor Head OFV-511/-512
B.2 Side Exit Head OFV-C-102 and OFV-C-110
The side exit heads have a mirror and a side exit window. The laser beam can
thus be deflected by 90 degrees from the output direction from the fiber.
The following side exit heads are available for the sensor head OFV-511/-512:
•
OFV-C-102, 90° steering head, diameter 5 mm
•
OFV-C-110, 90° steering head, diameter 10 mm
Assembly
The side exit head OFV-C-110 with mini sensor mounted is shown in
. The assembly of the side exit head OFV-C-102 are made in the
same way. To mount the side exit head, proceed as follows:
1. Place the mini sensor 4 into the side exit head as far as the shoulder of
the connector for the mini sensor is positioned at the edge of the opening
5.
The distance from the edge of the opening 5 to the exit window 1 is
about 130 mm.
2. Fix the mini sensor by gently tightening the stud screw 6.
3. Fix the holding device provided onto the grip 7 of the side exit head.
4. Position the side exit head at the measurement location.
5. You can position the exit window 1 subsequently by removing the knurled
screw 3, turning the probe 2 and fixing the knurled screw again.
6. Focus the laser beam by turning the focusing ring of the mini sensor 4
through the opening on the side exit head.
Figure B.3: Side exit head OFV-C-110 with mini sensor mounted
1
3
4
6
7
2
5