Resetting the Factory Default Values
1. Press to turn the Circulator Off.
2. Place the Mains Switch on the rear of the unit in the Off position.
3. Return the Mains Switch to the On position while pressing and
holding either the or button.
The factory default values are:
Operational Parameter
Factory Default Value
Temperature Set Point
20ºC/68ºF
Temperature Scale
ºC
Timer
00:00
High Limit
99.9ºC/211.8ºF
Calibration Offset
0.0ºC
Preset 1
Set Point: 37°C
Timer: 0
Preset 2
Set Point: 44°C
Timer: 0
Preset 3
Set Point: 70°C
Timer: 0
Loss of Power Restart
In the event that electrical power is lost while your Water Bath is in use,
it will begin operating automatically at the set point temperature once
power is restored. The timer will begin counting up from the point at
which power was restored. POWER FAIL will appear on the display to
alert you that there was a power disruption. To clear the message, turn
the bath Off and then back On again by pressing .
Displaying the Firmware Version
The version number for the Water Bath’s firmware is accessed by
pressing until the word VERSION appears on the LCD. The version
number will be displayed on the lower readout; this is a read-only value.
Press or allow the display to timeout (~10 seconds) to exit the Menu
function.