110-031 PSC/EN

9

Setting the High Limit Value

This menu item allows you to enter a high temperature alarm value, and 

can establish a temperature above which a set point temperature cannot 

be entered. If you try to increase the set point to a temperature higher 

than the High Limit value, further increases will be blocked and the words 

HIGH LIMIT on the LCD will flash.

1.  Press     until the words HIGH LIMIT appear on the LCD; the current 

High Limit value will appear on the lower readout.

2.  Use the      and      buttons to set the high limit value to the desired 

temperature; this value may be set as high as 105ºC (221ºF).

3.  Press      or allow the display to timeout (~10 seconds) to accept the 

displayed High Limit value. The words HIGH LIMIT will no longer be 

lit.

1.  Press     until PRESET appears on the lower readout.

2.  Use the     and     buttons to adjust the set point displayed on the 

lower readout to the desired temperature. 

3.  Press      until the word TIMER is displayed on the LCD; the two left 

hand digits in the lower readout will begin flashing.

4.  Use the     and     buttons to set the time duration in hours; this value 

may be set anywhere from 00 to 99 hours.

5.  Press     again; the two right hand digits in the lower readout will 

begin flashing.

6.  Use the     and     buttons to set the time duration in minutes; this 

value may be set anywhere from 00 to 59 minutes.

7.  Press      or allow the display to timeout (~10 seconds) to accept the 

displayed Time Duration value.

8.  Press and hold the Preset button (    ,     or     ) you wish to associate 

with the set point and time optional timer values you entered for ten 

seconds to accept. The unit will beep to indicate the Preset has been 

accepted.

Setting the Temperature Presets

NOTE

: Steps 3 through 7 below are optional.

!

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Содержание WBE10A11B

Страница 1: ...110 031 PSC EN 1 110 031 PSC EN 25 February 2014 Operator s Manual General Purpose Water Baths www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Страница 2: ...h ON 10 Adjusting the Set Point Temperature 10 Setting and Using the Timer 11 Using the Temperature Presets 11 Entering a Calibration Offset 12 Resetting the Factory Default Values 13 Loss of Power Re...

Страница 3: ...according to the directions in this manual and common safety procedures your Water Bath should provide safe and reliable temperature control Please ensure that all individuals involved in the install...

Страница 4: ...luid in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptable levels Always refill the reservoir using the same bath fluid type that is already in the res...

Страница 5: ...packing Your Water Bath Your water bath is shipped in a special carton Retain the carton and all packing materials until the unit is completely assembled and working properly Set up and run the unit i...

Страница 6: ...ID is being displayed on the lower readout Temperature Unit Symbol Shows temperature unit in use when lit Upper Readout Displays actual bath temperature Lower Readout Displays set point temperature d...

Страница 7: ...ss than 2 inches 5 08 cm from the bottom of the unit Operation of the bath without fluid will not damage the heater but will cause permanent discoloration of the tank and will not provide accurate tem...

Страница 8: ...ure to follow your organization s procedures and practices regarding the safe lifting and relocation of heavy objects WARNING To avoid the potential for burns allow the Water Bath to cool completely b...

Страница 9: ...he unit The extension cord must not cause more than a 10 drop in voltage to the unit Controller Setup Power On 1 Place the Mains Switch on the rear of the unit in the On position The three decimal poi...

Страница 10: ...er readout 2 Use the and buttons to adjust the set point displayed on the lower readout to the desired temperature 3 Press until the word TIMER is displayed on the LCD the two left hand digits in the...

Страница 11: ...bath is in the Standby mode 2 Press The actual bath temperature is displayed on the upper readout and the set point temperature is displayed on the lower readout Both values are displayed to one decim...

Страница 12: ...ded to inform you about how much time has elapsed since the timer reached 00 00 To clear the Timer Value place the unit in Standby mode and press to return from Standby mode Using the Temperature Pres...

Страница 13: ...T on the LCD will flash 1 Place the Mains Switch on the rear of the unit in the Off position 2 Press and hold while returning the Mains Switch to the On position 3 When the unit enters Standby the thr...

Страница 14: ...event that electrical power is lost while your Water Bath is in use it will begin operating automatically at the set point temperature once power is restored The timer will begin counting up from the...

Страница 15: ...sired set point temperature is displayed on lower readout Press or allow the display to timeout to accept C or F on lower readout Informational Indicates that the temperature unit is being changed Pre...

Страница 16: ...tion your Water Bath incorporates redundant over temperature safeties These over temperature low liquid level safeties automatically remove power from the heater in the event that the bath temperature...

Страница 17: ...uid as required Insufficient heating Improper line voltage Recent change in set point or heat load Bath cover not in place Check that line voltage meets specifications Allow sufficient time for bath t...

Страница 18: ...liter 10 6 x 11 6 x 6 26 9 x 29 5 x 15 2 cm 1000 WB10 1 8 6 A WB10 2 4 5 A 20 liter 9 5 x 17 5 x 5 24 1 x 44 5 x 15 2 cm 1400 WB20 1 12 0 A WB20 2 6 0 A 28 liter 9 5 x 17 x 8 24 1 x 43 2 x 20 3 cm 14...

Страница 19: ...distilled H2 0 5 C 0 92 3 85 5 to 90 C 10 to 107 C polycool MIX 25 50 50 mix with distilled H20 20 C 0 78 3 26 25 to 100 C 30 to 115 C polycool MIX 25 30 70 mix with distilled H20 0 C 0 89 3 72 0 to...

Страница 20: ...have to be frequently replenished Caution must be taken to stay well below the fluid s flashpoint Application Notes For optimum results maintain fluid level throughout the operating period adding flui...

Страница 21: ...ycling It is also your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect the persons involved in the disposal and recyclin...

Страница 22: ...t should be shipped transportation charges pre paid in either its original packaging or packaging affording an equal degree of protection to the indicated service center To insure prompt handling the...

Отзывы: