110-827
9
Troubleshooting Chart
Problem Cause
Corrective
Action
No power to unit
Electrical power
disconnected
Power Switch / Circuit
Breaker in Off position
Compressor overload switch
has tripped
Check that electrical cord is secure and plugged
into an operating electrical outlet.
Place Power Switch / Circuit Breaker in On
position.
Allow approximately 10 minutes for unit to restart.
No or insufficient
cooling
Blocked air flow
High ambient temperature
Improper line voltage and/or
frequency
Check air vents for blockage.
Check that air flow around unit is not restricted
and that room temperature does not exceed 95°F
/ 35°C.
Verify that Mains voltage during start cycle is
within 10% of rated voltage.
Gradual loss of cooling
Blocked air flow
Check air vents for blockage.
Compressor shakes,
stalls, or continually
restarts
Compressor cycling or
overload
Check air vents for blockage.
Check that air flow around unit is not restricted
and that room temperature does not exceed 95°F
/ 35°C.
Verify that Mains voltage during start cycle is
within 10% of rated voltage.
Check compressor fan for operation (listen for fan
noise, check airflow through unit).
Technical Information
Performance Specifications
Working Temperature:
Fixed at -60°C / -76°F
Reservoir Capacity:
1.9 liters
Tank:
Insulated stainless steel
Bath Work Area:
4.8
∅
x 6.6 inches deep / 12.2
∅
x 16.8 cm deep
Overall Dimensions (L x W x H): 22.75 x 11 x 11 inches / 57.8 x 27.9 x 27.9 cm
Electrical Requirements:
60 Hz: 120V, 60Hz, 5.0A
50 Hz: 240V, 50Hz, 2.5A
Specifications subject to change without notice.
Environmental Conditions
Indoor use only
Maximum Altitude:
2000 meter
Operating Ambient:
5° to 35°C (41° to 95°F)
Relative Humidity:
80%, non-condensing
Installation
Category:
II
Pollution
Degree:
2