background image

110-827 

6  

Installation 

Your Histology Bath is simple to setup and install. The only tool required is a container for adding a 
suitable fluid to the bath reservoir. 

General Site Requirements 

 

WARNING: The Histology Bath uses isopentane as the bath fluid. Due to the flammability of 
this liquid, always use the Histology Bath under a fume hood.  

Locate your Histology Bath on a level surface in an area that is free from drafts and wide ambient 
temperature variations, such as near heater or air conditioning vents. Do not place it where there 
are corrosive fumes, excessive moisture, or in excessively dusty areas.  

Position the unit to allow for unobstructed air flow through the side and rear vents. A space of at 
least 4 inches (10.2 cm) should be provided behind the unit.  

Avoid voltage drops by using properly grounded power outlets wired with 14 gauge or larger 
diameter wire and if possible, be close to the power distribution panel. The use of extension cords is 
not recommended; this will reduce the potential for problems caused by low line voltage. 

Adding Liquid to the Bath Reservoir 

 

WARNING: Fill reservoir with fluid before applying power to the unit. 

 

 

WARNING: Read the safety data sheet for the bath fluid being used carefully before filling 
reservoir. 

The Histology Bath is designed for use with isopentane (also known as 2-methlybutane).  

Fill bath so that the liquid level is approximately 1 inch (2.54 cm) from the top for best refrigeration 
performance. Be sure to allow for the fluid displacement that will result when the immersion basket 
and samples are placed in the reservoir.   

 

WARNING: Always drain all fluid from the reservoir before moving or lifting your Histology 
Bath. Be sure to follow your organization’s procedures and practices regarding the safe lifting 
and relocation of heavy objects. 

 

 

WARNING: To avoid the potential for burns, allow the Histology Bath to return to ambient 
temperature before cleaning or performing any maintenance.  

Electrical Power 

 

WARNING: The Histology Bath’s power cord must be connected to a properly grounded 
electrical receptacle. Make certain that this electrical outlet is the same voltage and frequency 
as your unit. The correct voltage and frequency for your Histology Bath are indicated on the 
identification label on the rear of the unit. 

 

 

CAUTION: The use of an extension cord is not recommended. If one is necessary, it must be 
properly grounded and capable of handling the total wattage of the unit. The extension cord 
must not cause more than a 10% drop in voltage to the unit.  

Plug the Histology Bath’s power cord into a properly grounded electrical outlet of the correct 
voltage and frequency. 

 

Содержание Histology Bath

Страница 1: ...Operator s Manual Histology Bath 110 827 09 12 11...

Страница 2: ...allation 6 General Site Requirements 6 Adding Liquid to the Reservoir 6 Electrical Power 6 Normal Operation 7 Routine Maintenance Troubleshooting 8 Draining the Reservoir 8 Cleaning the Bath 8 Air Ven...

Страница 3: ...ide safe operation Please ensure that all individuals involved in the installation operation or maintenance of this Histology Bath read this manual thoroughly prior to working with the unit This symbo...

Страница 4: ...d in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptable levels Always refill the reservoir using the same bath fluid type that is already in the reserv...

Страница 5: ...nty repaired but not replaced If the unit is damaged or does not operate properly contact the transportation company file a damage claim then contact the company where your unit was purchased WARNING...

Страница 6: ...110 827 5 Components Controls Power Switch Circuit Breaker Reservoir Cover Drain Valve Drain Port IEC Power Connection Reusable Air Filter behind grille...

Страница 7: ...arefully before filling reservoir The Histology Bath is designed for use with isopentane also known as 2 methlybutane Fill bath so that the liquid level is approximately 1 inch 2 54 cm from the top fo...

Страница 8: ...on the front of the Histology Bath in the On position The compressor will start up and the unit will begin cooling It will take about one hour for the unit to achieve a 60 C 76 F working temperature O...

Страница 9: ...l scouring pads are acceptable The entire unit is housed in a tough well insulated stainless steel casing that is corrosion and chemical resistant Air Vents The unit draws air through the rear and exh...

Страница 10: ...ockage Compressor shakes stalls or continually restarts Compressor cycling or overload Check air vents for blockage Check that air flow around unit is not restricted and that room temperature does not...

Страница 11: ...nation so as to protect the persons involved in the disposal and recycling of the equipment from health hazards By doing so you will help to conserve natural and environmental resources and you will e...

Страница 12: ...the outside of the package A detailed explanation of the defect should be enclosed with the item The warranty shall not apply if the defect or malfunction was caused by accident neglect unreasonable u...

Отзывы: