background image

Operator’s Manual

 

Histology Bath 

 

 

 
 
 

      

110-827

09.12.11 

Содержание Histology Bath

Страница 1: ...Operator s Manual Histology Bath 110 827 09 12 11...

Страница 2: ...allation 6 General Site Requirements 6 Adding Liquid to the Reservoir 6 Electrical Power 6 Normal Operation 7 Routine Maintenance Troubleshooting 8 Draining the Reservoir 8 Cleaning the Bath 8 Air Ven...

Страница 3: ...ide safe operation Please ensure that all individuals involved in the installation operation or maintenance of this Histology Bath read this manual thoroughly prior to working with the unit This symbo...

Страница 4: ...d in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptable levels Always refill the reservoir using the same bath fluid type that is already in the reserv...

Страница 5: ...nty repaired but not replaced If the unit is damaged or does not operate properly contact the transportation company file a damage claim then contact the company where your unit was purchased WARNING...

Страница 6: ...110 827 5 Components Controls Power Switch Circuit Breaker Reservoir Cover Drain Valve Drain Port IEC Power Connection Reusable Air Filter behind grille...

Страница 7: ...arefully before filling reservoir The Histology Bath is designed for use with isopentane also known as 2 methlybutane Fill bath so that the liquid level is approximately 1 inch 2 54 cm from the top fo...

Страница 8: ...on the front of the Histology Bath in the On position The compressor will start up and the unit will begin cooling It will take about one hour for the unit to achieve a 60 C 76 F working temperature O...

Страница 9: ...l scouring pads are acceptable The entire unit is housed in a tough well insulated stainless steel casing that is corrosion and chemical resistant Air Vents The unit draws air through the rear and exh...

Страница 10: ...ockage Compressor shakes stalls or continually restarts Compressor cycling or overload Check air vents for blockage Check that air flow around unit is not restricted and that room temperature does not...

Страница 11: ...nation so as to protect the persons involved in the disposal and recycling of the equipment from health hazards By doing so you will help to conserve natural and environmental resources and you will e...

Страница 12: ...the outside of the package A detailed explanation of the defect should be enclosed with the item The warranty shall not apply if the defect or malfunction was caused by accident neglect unreasonable u...

Отзывы: