3.2 Device operation 

 
3.2.1 Turning on the device turn, testing and calibration 
3.2.1.1 To turn the device ON, press the ON/OFF button 8 (Figure 2). The device enters the 

test mode after it is turned ON or after the  batteries are replaced. While the device is in this mode, 
Red-LED and Yellow-LED are flashing in turn. After completion of the test mode the device can go 
either to the calibration mode (if batteries were replaced), or to the mode of detection of exceeding 
the  preset  thresholds.  The  device  is  ready  for  operation,  when  Red-LED  and  Yellow-LED  are 
switched OFF and Green-LED is switched ON.

 

3.2.1.2  While  in  the  calibration  mode,  Red-LED  and  Yellow-LED  are  flashing 

simultaneously. The calibration of TSM is performed in this mode. To calibrate the TSM, mount 
the filter 14 (Figure 2) (included in SET-O) to the input connecting pipe 2 and turn its cover 16 to 
the  “Open”  position. After  that  enter  the  calibration  mode  by  replacing  the  batteries.  The 
calibration will be completed  in no more than 2 min. The calibration completion is indicated by 
Red-LED and  Yellow-LED  switching  OFF.  After  the  calibration  completion,  turn  the filter  cover 
16 to the “Close” position, remove the filter 14 and bring it into the right position in the SET-O. 
The device automatically enters into the mode of detection of exceeding the preset thresholds 
of the gamma radiation DER and concentrations of toxic substance vapors.

 

 

Caution!  If  the  calibration  process  was  interrupted  (the  device  was  turned  OFF 

before  the  simultaneous  flashing  of  the  Yellow-LED  and  the  Red-LED  stopped),  then  in 
the  mode  of  detection  of  exceeding  of  the  set  thresholds  of  gamma  radiation  and 
concentrations  of  toxic  substance  vapors  the  Red-LED  will  be  flashing  with  10  s 
intervals.  In  this  case  to  ensure  detection  thresholds,  perform  calibration  until  both  the 
R-LED and the Y-LED are off.

 

3.2.2 Device operation in the mode of detection of exceeding the preset thresholds of the 

gamma radiation DER and concentrations of toxic substances vapors 

3.2.2.1 While in the mode of detection of exceeding the preset thresholds of the gamma 

radiation  DER  and  concentrations  of  toxic  substances  vapors,  the  device  can  be  used  for  the 
simultaneous detection of toxic substance vapors (TSV) in air and the detection and location of 
gamma radiation sources (GRS). 

The device operates with the parameter values set by the manufacturer (3.2.3.1). To use 

other parameters, it is necessary to set the other values (3.2.4). 

The device automatically enters the mode of detection of exceeding the preset thresholds 

of GRS and TSV after the completion of the calibration mode. 

3.2.2.2 Detection of GRS and TSV 
To  achieve  the  maximum  sensitivity  during  the  GRS  detection,  the  device  should  be 

placed  in  such  a  position  that  the  calibration  direction  indicated  in  Figure  2  coincides  with  a 
direction to a scanned object. 

The  detection  efficiency  is  dependent  on  the  proximity  of  the  device  to  an  examined 

object (thing, car, person, etc) and velocity of its movement along the object. 

To achieve the maximum sensitivity during the TSV detection, the input connecting pipe 

of the toxic substance module should be completely opened. 

3.2.2.3 Location of gamma radiation sources 
When a gamma radiation source is detected, it can be located.  
To  locate  the  source,  the  device  should  be  at  no  more  than  10  cm  distant  from  the 

scanned  object.  The  velocity  of  the  device  movements  along  the  scanned  object  should  not 
exceed  1cm/s.  As  the  device  moves  closer  to  the  gamma  radiation  source,  a  rate  at  which 
Yellow LED is flashing and audio tone is repeated will increase.  

When  the  maximum  rate  of  audio  signals  is  achieved,  a  continuous  audio  tone  is 

produced. The source can not be located better. 

Содержание PM2010

Страница 1: ...ALARMING DEVICE FOR DETECTION OF GAMMA RADIATION AND CHEMICAL WEAPONS VAPORS PM2010 OPERATING MANUAL ...

Страница 2: ...3 1 Preparation for use 13 3 2 Device operation 14 4 MAINTENANCE 16 5 TROUBLESHOOTING 19 6 STORAGE AND SHIPPING 20 7 UTILIZATION 21 8 WARRANTY 22 9 PACKING CERTIFICATE 23 10 CERTIFICATE OF PUTTING INTO SERVICE 24 11 ACCEPTANCE CERTIFICATE 25 12 WARRANTY CERTIFICATE 26 13 CLAIMS 27 APPENDIX A Copy of certificate toxic substance detection module tsm 28 APPENDIX B Beta radiation source copy of passpo...

Страница 3: ...nance as well as some other information necessary for the proper operation of the device and full realization of its possibilities The name of the device in the documentation is written as follows Alarming device for detection of gamma radiation and chemical weapons vapors PM2010 Type 100345122 042 2004 During manufacturing of the device some changes may be introduced in its electrical scheme cons...

Страница 4: ...ration history is stored in its non volatile memory and can be transferred to a personal computer PC via an infrared IR interface of data transmission using IrDA interface The device may be operated both indoor and outdoor by a wide range users involved in detecting of ionizing radiation sources and vapors of toxic substances The device may be used by staff of emergency services by border guards c...

Страница 5: ...operating manual Appendix B Certificate of toxic substance module 1 piece included in operating manual Appendix A Battery Energizer L91BP 2 AA 1 2 pieces IR adapter ACT IR220L or IR 210B 2 1 piece Set of spare parts and tools SET O Filter 1 piece Rod 1 piece Screwdriver 1 piece Filtering tissue element 4 pieces Filtering element 4 pieces Bottle with filter filler 1 piece Coarse calico strip 0 4 m ...

Страница 6: ...the set thresholds are exceeded and record in non violate memory the events of set DER threshold exceeding and history of DER registration in case of supply voltage changes from 2 5 to 3 2 V 2 3 9 Threshold of detection with a toxic substance module at contents higher than threshold values at normal climatic conditions NCC temperature 20 5 o C atmospheric pressure 101 3 5 4 17 3 kPa relative humid...

Страница 7: ...5 mm repeated shocks with acceleration of 100 m s2 duration of shock pulse of 2 50 ms at shock rate of 60 180 shocks per minute 2 3 18 Device is resistant to RF electrical and magnetic fields under IEC 61000 4 3 1995 in the frequency range of 80 to 1000 MHz in the frequency range of 800 to 960 MHz and of 1 4 to 2 0 GHz generated by digital mobile phones 10 V m severity degree 3 30 V m severity deg...

Страница 8: ...he registered concentration value of toxic substance vapors with the set threshold value of concentration level The detection modules are operated by a microprocessor controller with a built in analogue digital converter ADC The device is powered by two AA batteries The nominal battery voltage is 1 5 V 2 4 2 A block diagram of the device is presented in Figure 1 The device consists of gamma radiat...

Страница 9: ... 2 with a cap 15 is located on the top cover of the device When the device is operated the cap is taken off the connecting pipe 2 and fixed adjacent to the connecting pipe at a plug 1 During the device calibration the filter 14 involved in the set of spare parts and tools SET O is screwed on the connecting pipe 2 The Red LED 4 Yellow LED 5 Green LED 6 ON OFF button 8 window 7 of the IR interface a...

Страница 10: ... window with the device serial number 4 Yellow LED 5 Red LED 6 Green LED 7 input window of the IR interface 8 ON OFF button 9 GMM geometric center 10 cap of the output connecting pipe 11 housing with batteries 12 window for the audio alarming device 13 clip 14 filter 15 cap 16 filter cover Figure 2 ...

Страница 11: ...itions While in this mode the device produces light and audible signals as follows Chemical alarm Red LED is flashing when the threshold value of the TSM is exceeded Radiation alarm Yellow LED is flashing when the preset DER threshold is exceeded The light alarms are accompanied by audible alarms if these are enabled in the mode of communication with PC The device provides for a possibility to rec...

Страница 12: ...on the device The Radiation danger sign with the source type and its activity is marked on the bottom side of the device housing A seal that prevents an unauthorized opening of the device is placed in the faceplate of the device in the screw groove 2 4 6 Packing The device is put into a pasteboard box together with its documents and SET ...

Страница 13: ...crew cap 15 on the connecting pipe at a plug 1 and put cap 10 of the output connecting pipe into SET O 3 1 3 6 When starting the operation after the long storage of the device it is required to clean the gas system and the ionizing chamber in according to clause 4 4 of the present Manual 3 1 3 7 Calibrate the device according to clause 3 2 1 2 After calibration completion the device is ready for o...

Страница 14: ... to ensure detection thresholds perform calibration until both the R LED and the Y LED are off 3 2 2 Device operation in the mode of detection of exceeding the preset thresholds of the gamma radiation DER and concentrations of toxic substances vapors 3 2 2 1 While in the mode of detection of exceeding the preset thresholds of the gamma radiation DER and concentrations of toxic substances vapors th...

Страница 15: ...solution of a monitor IrDA Connect the cable of the IR adapter to a communication port of the personal computer or use the PC with a built in IR adapter To install the application program use the CD with software included in the delivery kit The Help file describes how to run the application program To communicate with PC turn the device ON and place it for 15 s near the IR adapter at a distance o...

Страница 16: ...2 After the beta radiation source service life indicated in source passport Appendix B is expired it should be replaced by a new one The radioactive source replacement should be performed by the specially trained personnel of the manufacturer personnel of A group 4 2 3 The utilization of used sources should follow a determinate order observing the requirements of handling radioactive materials 4 3...

Страница 17: ...To blow out the gas system it is necessary to turn the device ON the filter cover open and hold it ON for one hour False chemical alarms may occur After an hour turn the device OFF and close the filter cover CAUTION When the maintenance of the gas system and ionizing chamber is performed alcohol drops and tissue fibers should not be left in the ionizing chamber and connecting pipe 4 5 Replacement ...

Страница 18: ...SET O 6 washer 7 spring 8 nut 9 cover Figure 3 4 6 Device deactivation 4 6 1 The maintenance involves the external examination dust and dirt removal and decontamination in case when radioactive substances are fallen on the device housing Decontamination is carried out by wiping the device housing with a soft tissue moistens with ethyl alcohol ...

Страница 19: ... of TSM Cap of the ionizing chamber is not removed Blocking of the gas system and ionizing chamber Imitator cover is not removed TSM module is failed Remove the cap Clean the gas system and ionizing chamber according to clause 4 4 Remove the cover Send device to the manufacturer 4 Red LED and Yellow LED are flashing in turn for more than 2 minutes One of the modules is failed Send device to the ma...

Страница 20: ...ay cause corrosion 6 2 Devices in package may be shipped by any kinds of enclosed transport at the air temperature from 20 to 50 C Devices in package should be fastened in a vehicle They must be arranged and fastened in transport so as to ensure their stable position and to avoid shocks to each other and the walls of a vehicle When carried by see devices in package should be placed in hermetic pla...

Страница 21: ...ices of the group 3 The utilization of the used source should be performed according to the requirements of regulations and safety measures for the radioactive source handling should be observed 7 2 Precious metal data contented in the device are not quoted as their weight is not higher than the values stated in the Sate Standard GOST 2 608 78 ...

Страница 22: ...period is 6 months since the date of manufacturer Quality Control Department accepting the device 8 4 Warranty and after warranty repairs are performed by the manufacturer or by a company authorized by the manufacturer 8 5 Warranty does not cover the devices subjected to the user s service the device was tampered opened by the user with mechanical damages if the requirements of operation and stora...

Страница 23: ...___ serial number packed ______________________________________________________________ name of manufacturer According to the requirements of design documentation __________________________ ________________________ ______________________ position signature family name ____________________________________ year month date stamp ...

Страница 24: ...or detection of gamma radiation and chemical weapons vapors PM2010 Serial number ______________ is put into service ______________________ date of putting into service ___ _________________ _________________________ signature and name of the responsible person ...

Страница 25: ... vapors PM2010 Serial number ___________________ is manufactured to meet the requirements of State Standards and the Design documentation and accepted for operation Head of Quality Control Department stamp place _____________________ ________________ signature name ___ ________________ _____________ ...

Страница 26: ...0 81 Fax 375 17 260 23 56 E mail polimaster polimaster com E mail plant polimaster com Address 112 M Bogdanovich str Minsk 220040 Republic of Belarus Address 51 Skoriny str Minsk 220141 Republic of Belarus Date of sale ___ ________ _____ Seller ______________ signature Seller stamp _________________ The warranty after warranty repair is made ___ ____________ ______ The warranty period is prolonged...

Страница 27: ...27 13 CLAIMS ...

Страница 28: ...anophosphorus substances sarin soman V gases and arsenic containing substances lewisite in air A 3 DELIVERY KIT A 3 1 TSM delivery kit includes Toxic substance detection module 1 piece Quality certificate passport of radioactive source Appendix 2 of the present Manual 1 piece SET O set of spare parts and tools 1 set batch of TSM delivered to one address is supplied with 3 operating manuals per 10 ...

Страница 29: ...by the source external radiation nA ______________ B 2 4 Relative error of the ionization current value at 0 95 confidence level 15 B 2 5 The radioactive contamination of the source external surface does not exceed the permissible value An activity of radioactive substances taken from the source surface by the smear method of determination of the radioactive contamination level is no more than 5nC...

Страница 30: ...erature from 50 o C to 50 o C and a relative humidity up to 98 at 40 o C B 6 3 Do not remove seals and open package during the source transportation and storage B 6 4 Sources should not be stored in rooms provided with a closed ventilation system as well as in the same room with inflammable explosive substances and aggressive chemicals B 7 OPERATION GUIDELINES B 7 1 When operating the source requi...

Страница 31: ...31 ...

Отзывы: