 3 USE OF THE DEVICE 
 

3.1 Preparation for use 

3.1.1 General information 
3.1.1.1 When  buying  the  device  please  check  the  delivery  kit  according  to  clause  2.2.1 

and perform the device operation check up according to clause 3.1.4. 

The device should not be subjected to severe mechanical shocks, any strong chemicals 

and solvents and open fire sources. 

 
3.1.2 Safety precautions 
3.1.2.1  During  the  device  assembling,  calibration,  repair  and  maintenance  when 

radioactive  sources  are  used,  all  the  requirements  and  regulations  ensuring  the  radiation  and 
sanitary safety should be observed.  

3.1.2.2  The  beta  radiation  source  IBIRRZN-63-IIa,  7016-056-41805307-2002  which  is 

included  in  the  ionization  chamber  contains  the 

63

Ni  radionuclide  of  0.6  GBq  activity.  This 

source is located inside of the ionizing chamber in the device housing. 

3.1.2.3  The  beta  radiation  DER  does  not  exceed  the  natural  background  level  at  any 

point of the outer surface of the package with the device inside. The device design excludes the 
possibility of the access to the source during device operation. Therefore it is not necessary to 
take any additional safety precautions actions for radiation security. 

 
3.1.3 Preparation for operation 
3.1.3.1 Before using the device, please read this operating manual carefully.  
3.1.3.2 Take the device from the package. 
3.1.3.3 Insert the batteries into the battery compartment 11 (Figure 2). 
3.1.3.4  Remove  cap  15  from  the  connecting  pipe  of  the  ionizing  chamber  2  and  the 

output connecting pipe 10. 

3.1.3.5  Screw  cap  15 on  the  connecting  pipe  at  a  plug  1,  and  put  cap  10  of the output 

connecting pipe into SET-O. 

3.1.3.6 When starting the operation after the long storage of the device, it is required to clean 

the gas system and the ionizing chamber in according to clause 4.4 of the present Manual. 

3.1.3.7 Calibrate the device according to clause 3.2.1.2. After calibration completion the 

device is ready for operation  

 
3.1.4 The device operation checking 
3.1.4.1 After device calibration according to clause 3.2.1 it is necessary to check up the 

device  operation.  After  calibration  completion,  the  Green-LED  should  light  constantly  and  a 
sound of the working micropump should be heard.  

To  check  up  the  chemical  channel  operation,  bring  the  imitator  from  the  SET-O  (set  of 

the  spare  parts  and  tools)  to  the  input  of  the  ionizing  chamber.  The  light  alarm  (Red-LED) 
should flash and audio alarm will be heard. Remove the imitator. The light and audio alarms will 
be switched OFF in no more than 30 s. 

Caution! When checking up the chemical channel operation with the imitator from 

the  SET-O  in  enclosed  space  it  is  necessary  to  provide  the  blowout  from  the  output 
connecting  pipe  outdoors  (exhaust  ventilation).  Before  the  iterative  checking  up  it  is 
necessary to blow out the gas system through the TSM filter for 10-20 minutes

 

To  check  up  the  radiation  channel,  bring  a  radioactive  source  (its  energy  range  is  from 

0.06 to 3.0 MeV)  close to the GMM geometrical center 13 (Figure 2). The light alarm (Yellow-
LED) will flash and audio alarm will be heard. Remove the source. The light and audio alarms 
will be switched OFF for no more than 30 s. 

To turn the device OFF, press the button 8. The Green-LED 6 and the micropump should 

be switched OFF. 

Содержание PM2010

Страница 1: ...ALARMING DEVICE FOR DETECTION OF GAMMA RADIATION AND CHEMICAL WEAPONS VAPORS PM2010 OPERATING MANUAL ...

Страница 2: ...3 1 Preparation for use 13 3 2 Device operation 14 4 MAINTENANCE 16 5 TROUBLESHOOTING 19 6 STORAGE AND SHIPPING 20 7 UTILIZATION 21 8 WARRANTY 22 9 PACKING CERTIFICATE 23 10 CERTIFICATE OF PUTTING INTO SERVICE 24 11 ACCEPTANCE CERTIFICATE 25 12 WARRANTY CERTIFICATE 26 13 CLAIMS 27 APPENDIX A Copy of certificate toxic substance detection module tsm 28 APPENDIX B Beta radiation source copy of passpo...

Страница 3: ...nance as well as some other information necessary for the proper operation of the device and full realization of its possibilities The name of the device in the documentation is written as follows Alarming device for detection of gamma radiation and chemical weapons vapors PM2010 Type 100345122 042 2004 During manufacturing of the device some changes may be introduced in its electrical scheme cons...

Страница 4: ...ration history is stored in its non volatile memory and can be transferred to a personal computer PC via an infrared IR interface of data transmission using IrDA interface The device may be operated both indoor and outdoor by a wide range users involved in detecting of ionizing radiation sources and vapors of toxic substances The device may be used by staff of emergency services by border guards c...

Страница 5: ...operating manual Appendix B Certificate of toxic substance module 1 piece included in operating manual Appendix A Battery Energizer L91BP 2 AA 1 2 pieces IR adapter ACT IR220L or IR 210B 2 1 piece Set of spare parts and tools SET O Filter 1 piece Rod 1 piece Screwdriver 1 piece Filtering tissue element 4 pieces Filtering element 4 pieces Bottle with filter filler 1 piece Coarse calico strip 0 4 m ...

Страница 6: ...the set thresholds are exceeded and record in non violate memory the events of set DER threshold exceeding and history of DER registration in case of supply voltage changes from 2 5 to 3 2 V 2 3 9 Threshold of detection with a toxic substance module at contents higher than threshold values at normal climatic conditions NCC temperature 20 5 o C atmospheric pressure 101 3 5 4 17 3 kPa relative humid...

Страница 7: ...5 mm repeated shocks with acceleration of 100 m s2 duration of shock pulse of 2 50 ms at shock rate of 60 180 shocks per minute 2 3 18 Device is resistant to RF electrical and magnetic fields under IEC 61000 4 3 1995 in the frequency range of 80 to 1000 MHz in the frequency range of 800 to 960 MHz and of 1 4 to 2 0 GHz generated by digital mobile phones 10 V m severity degree 3 30 V m severity deg...

Страница 8: ...he registered concentration value of toxic substance vapors with the set threshold value of concentration level The detection modules are operated by a microprocessor controller with a built in analogue digital converter ADC The device is powered by two AA batteries The nominal battery voltage is 1 5 V 2 4 2 A block diagram of the device is presented in Figure 1 The device consists of gamma radiat...

Страница 9: ... 2 with a cap 15 is located on the top cover of the device When the device is operated the cap is taken off the connecting pipe 2 and fixed adjacent to the connecting pipe at a plug 1 During the device calibration the filter 14 involved in the set of spare parts and tools SET O is screwed on the connecting pipe 2 The Red LED 4 Yellow LED 5 Green LED 6 ON OFF button 8 window 7 of the IR interface a...

Страница 10: ... window with the device serial number 4 Yellow LED 5 Red LED 6 Green LED 7 input window of the IR interface 8 ON OFF button 9 GMM geometric center 10 cap of the output connecting pipe 11 housing with batteries 12 window for the audio alarming device 13 clip 14 filter 15 cap 16 filter cover Figure 2 ...

Страница 11: ...itions While in this mode the device produces light and audible signals as follows Chemical alarm Red LED is flashing when the threshold value of the TSM is exceeded Radiation alarm Yellow LED is flashing when the preset DER threshold is exceeded The light alarms are accompanied by audible alarms if these are enabled in the mode of communication with PC The device provides for a possibility to rec...

Страница 12: ...on the device The Radiation danger sign with the source type and its activity is marked on the bottom side of the device housing A seal that prevents an unauthorized opening of the device is placed in the faceplate of the device in the screw groove 2 4 6 Packing The device is put into a pasteboard box together with its documents and SET ...

Страница 13: ...crew cap 15 on the connecting pipe at a plug 1 and put cap 10 of the output connecting pipe into SET O 3 1 3 6 When starting the operation after the long storage of the device it is required to clean the gas system and the ionizing chamber in according to clause 4 4 of the present Manual 3 1 3 7 Calibrate the device according to clause 3 2 1 2 After calibration completion the device is ready for o...

Страница 14: ... to ensure detection thresholds perform calibration until both the R LED and the Y LED are off 3 2 2 Device operation in the mode of detection of exceeding the preset thresholds of the gamma radiation DER and concentrations of toxic substances vapors 3 2 2 1 While in the mode of detection of exceeding the preset thresholds of the gamma radiation DER and concentrations of toxic substances vapors th...

Страница 15: ...solution of a monitor IrDA Connect the cable of the IR adapter to a communication port of the personal computer or use the PC with a built in IR adapter To install the application program use the CD with software included in the delivery kit The Help file describes how to run the application program To communicate with PC turn the device ON and place it for 15 s near the IR adapter at a distance o...

Страница 16: ...2 After the beta radiation source service life indicated in source passport Appendix B is expired it should be replaced by a new one The radioactive source replacement should be performed by the specially trained personnel of the manufacturer personnel of A group 4 2 3 The utilization of used sources should follow a determinate order observing the requirements of handling radioactive materials 4 3...

Страница 17: ...To blow out the gas system it is necessary to turn the device ON the filter cover open and hold it ON for one hour False chemical alarms may occur After an hour turn the device OFF and close the filter cover CAUTION When the maintenance of the gas system and ionizing chamber is performed alcohol drops and tissue fibers should not be left in the ionizing chamber and connecting pipe 4 5 Replacement ...

Страница 18: ...SET O 6 washer 7 spring 8 nut 9 cover Figure 3 4 6 Device deactivation 4 6 1 The maintenance involves the external examination dust and dirt removal and decontamination in case when radioactive substances are fallen on the device housing Decontamination is carried out by wiping the device housing with a soft tissue moistens with ethyl alcohol ...

Страница 19: ... of TSM Cap of the ionizing chamber is not removed Blocking of the gas system and ionizing chamber Imitator cover is not removed TSM module is failed Remove the cap Clean the gas system and ionizing chamber according to clause 4 4 Remove the cover Send device to the manufacturer 4 Red LED and Yellow LED are flashing in turn for more than 2 minutes One of the modules is failed Send device to the ma...

Страница 20: ...ay cause corrosion 6 2 Devices in package may be shipped by any kinds of enclosed transport at the air temperature from 20 to 50 C Devices in package should be fastened in a vehicle They must be arranged and fastened in transport so as to ensure their stable position and to avoid shocks to each other and the walls of a vehicle When carried by see devices in package should be placed in hermetic pla...

Страница 21: ...ices of the group 3 The utilization of the used source should be performed according to the requirements of regulations and safety measures for the radioactive source handling should be observed 7 2 Precious metal data contented in the device are not quoted as their weight is not higher than the values stated in the Sate Standard GOST 2 608 78 ...

Страница 22: ...period is 6 months since the date of manufacturer Quality Control Department accepting the device 8 4 Warranty and after warranty repairs are performed by the manufacturer or by a company authorized by the manufacturer 8 5 Warranty does not cover the devices subjected to the user s service the device was tampered opened by the user with mechanical damages if the requirements of operation and stora...

Страница 23: ...___ serial number packed ______________________________________________________________ name of manufacturer According to the requirements of design documentation __________________________ ________________________ ______________________ position signature family name ____________________________________ year month date stamp ...

Страница 24: ...or detection of gamma radiation and chemical weapons vapors PM2010 Serial number ______________ is put into service ______________________ date of putting into service ___ _________________ _________________________ signature and name of the responsible person ...

Страница 25: ... vapors PM2010 Serial number ___________________ is manufactured to meet the requirements of State Standards and the Design documentation and accepted for operation Head of Quality Control Department stamp place _____________________ ________________ signature name ___ ________________ _____________ ...

Страница 26: ...0 81 Fax 375 17 260 23 56 E mail polimaster polimaster com E mail plant polimaster com Address 112 M Bogdanovich str Minsk 220040 Republic of Belarus Address 51 Skoriny str Minsk 220141 Republic of Belarus Date of sale ___ ________ _____ Seller ______________ signature Seller stamp _________________ The warranty after warranty repair is made ___ ____________ ______ The warranty period is prolonged...

Страница 27: ...27 13 CLAIMS ...

Страница 28: ...anophosphorus substances sarin soman V gases and arsenic containing substances lewisite in air A 3 DELIVERY KIT A 3 1 TSM delivery kit includes Toxic substance detection module 1 piece Quality certificate passport of radioactive source Appendix 2 of the present Manual 1 piece SET O set of spare parts and tools 1 set batch of TSM delivered to one address is supplied with 3 operating manuals per 10 ...

Страница 29: ...by the source external radiation nA ______________ B 2 4 Relative error of the ionization current value at 0 95 confidence level 15 B 2 5 The radioactive contamination of the source external surface does not exceed the permissible value An activity of radioactive substances taken from the source surface by the smear method of determination of the radioactive contamination level is no more than 5nC...

Страница 30: ...erature from 50 o C to 50 o C and a relative humidity up to 98 at 40 o C B 6 3 Do not remove seals and open package during the source transportation and storage B 6 4 Sources should not be stored in rooms provided with a closed ventilation system as well as in the same room with inflammable explosive substances and aggressive chemicals B 7 OPERATION GUIDELINES B 7 1 When operating the source requi...

Страница 31: ...31 ...

Отзывы: