20 

Sound pressure level at the distance of 
20 cm (for РМ1621М, РМ1621МА), no 
less 

  85 dB 

(А). 

21 

Power supply 

  1.5  (+  0.1;  minus  0.4)  V  (one  AA  type 

battery) 

22 

Period of continuous operation of the 
dosimeter from one battery, using LCD 
backlight not more than 20 s/day, 
audible alarm 

– not more than 20 s/day 

and at an average value of the 
measured DER up to 0.3  Sv/h: 

   

 

- at a temperature from 0 up to 60 C 

  12 months 

 

- at a temperature minus 40 C, at least: 

  6 months 

23 

Protection  de

gree  of  the  dosimeter’s 

case

 

   

IP67  

24 

The  dosimeter  is  proof  against  the 
action of: 

 

  -  temperature  of  an  ambient  air  from 

minus 20

 

up to plus 60  C with indication 

of  measurement  results  on  LCD;  
-  temperature  of  an  ambient  air  from 
minus  40  up  to  minus  20 C  without 
indication  of  measurement  results  on 
LCD but with recording of measurement 
results in a non-volatile memory; 
-  relative  humidity  of  an  ambient  air  up 
to 98 % at 35  C; 
-  atmosphere  pressure  from  84  up  to 
106,7 kPa 

25 

The dosimeter is proof against: 

 

  - sinusoidal vibration in the frequency 

range 5 - 35 Hz and bias amplitude for 
frequencies lower than the transition 
frequency 0.75 mm; 
-  shocks  with  acceleration  100  m/s

2

duration  of  shock  pulse  2-50  ms,  shock 
rate is 60 - 180 shocks per minute 

26 

The dosimeter meets drop test against a 
concrete surface from the height 

  0,7 m 

27 

The  dosimeter  is  proof  against  the 
action  of  static  and  variable  magnetic 
fields of strength  

  up to 400 

А/m 

28 

The dosimeter is proof against the action of radio frequency electromagnetic fields up 
to 30 V/m in  the frequency range  from  80  to 1000 MHz,  from  800 to 960 MHz, from 
1.4  to  2.5  HHz  of  strength  (in  electromagnetic  emission  envoronment  of  digital 
radiophone) 

 

- the dosimeter is proof against the action of electrostatic discharges, test level 3 (air 
discharge, 8 kV, contact discharge 6 kV) , performance criterion 

В; 

- dosimeter corresponds electromagnetic compatibility state standard of 
manufacturing country (EN 55022:1998 (class B)) requirements by the level of emitted 
radio noise 

29 

The dosimeter in a transport package is 
proof against the action of: 

  - temperature from minus 50 up to plus 

50  C; 
- humidity up to 100 % at 40  C; 
- shocks with acceleration 98 m/s

2

duration 16 ms; 
- vibrations with frequency 5-35 Hz and 
bias amplitude 0.75 mm 

Содержание PM1621

Страница 1: ...X RAY AND GAMMA RADIATION PERSONAL DOSIMETERS 1621 Models 1621 1621 1621 1621 OPERATING MANUAL...

Страница 2: ...and tools 27 5 3 Verification officers qualification requirements 27 5 4 Safety requirements 27 5 5 Verification conditions 28 5 6 Pre verification procedure 28 5 7 Verification procedure 28 5 7 1 Ext...

Страница 3: ...l includes the general description specifications of the dosimeter as well as some other information necessary for the proper operation of the dosimeter and a full realization of its possibilities Dur...

Страница 4: ...rs from 1621 model by extended DER measurement range search mode and integrated vibro and light alarm The dosimeters may be used independently or as a part of a system for everyday efficient and emerg...

Страница 5: ...supply element ANASONIC POWER LINE LR6 AA 2 5 or Power supply element Energizer L91BP 2 AA 3 5 1 1 1 1 software 1 CD 1 CD 1 CD 1 CD Operating manual 4 1 1 1 1 Cover 5 1 1 1 1 Packaging 1 1 1 1 1 Supp...

Страница 6: ...621 1 00 2 00 Sv h 1621 1621 DER analogue scale seven segments in a logarithmic gauge Number of indicated segments from left to right corresponds to the following DER threshold values on the LCD displ...

Страница 7: ...gy does not exceed values in presented in Table 1 2 when the Dosimeter is rotated in the horizontal plane Attachment B figure B 1 and values in presented in Table 1 3 when the Dosimeter is rotated in...

Страница 8: ...D DER or DE photon radiation measured value indication relative mean square error of average value of measurement result with 0 95 probability statistical error in percents dosimeter number blind dosi...

Страница 9: ...Hz and bias amplitude for frequencies lower than the transition frequency 0 75 mm shocks with acceleration 100 m s2 duration of shock pulse 2 50 ms shock rate is 60 180 shocks per minute 26 The dosim...

Страница 10: ...1 Overall dimensions no more not more than 87 72 39 mm 32 Reliability parameters average full operating time no less than 20000 h average service life no less than 10 years average time of recovery no...

Страница 11: ...ock diagram of the dosimeter A Geiger Muller tube with a filter for spatial energy formation of sensitivity which converts photon radiation quanta to electric pulses is used as a radiation detector Th...

Страница 12: ...DE measurement mode year of production in the dosimeter s number indication mode indication of IR communication channel switch on off in the PC communication mode 3 digital panel of indication of stat...

Страница 13: ...13 Figure 1 2 General overview of the dosimeter...

Страница 14: ...14 Figure 1 3 Demounting the clip from 1621 and 1621...

Страница 15: ...performs LCD testing all the segments and graphic symbols should be displayed for about 2 s and then disappear then the dosimeter should enter the measurement mode One minute after inserting the batte...

Страница 16: ...of DE accumulation time values in hours On On a display 4 output of the values of time remaining for staying at the working place Off Thresholds setting enable On DE reset On Audible alarm On Dosimet...

Страница 17: ...ible Calculation of time is performed relative to the second preset DE threshold the function may be switched on at the DE measurement mode configuration 2 4 5 Settings mode Auxiliary mode of settings...

Страница 18: ...III 1 G free HDD space printer and unit for operation with IrDA protocol for the exchange of information with the Dosimeters are necessary for comfort program running The program runs under OS Window...

Страница 19: ...d closer to GRS When limiting frequency is reached the dosimeter will emit continuous audio alarm At single pressing of LIGHT SET button the frequency of audio light and vibro alarms can be decreased...

Страница 20: ...the LCD will indicate the flashing symbol bat The battery is to be replaced 3 3 In case of the battery critical discharge approximately 0 9 V the dosimeter will turn into the DE measurement mode the...

Страница 21: ...21 Figure 2 2 Select the 1621 1621 dosimeter operating indication mode...

Страница 22: ...22 Figure 2 3 Setting the DE threshold values...

Страница 23: ...23 Figure 2 4 Setting the DER threshold values...

Страница 24: ...24 Figure 2 5 Indication of audio alarm state on off adjustment of sound volume of 1621 1621 dosimeter...

Страница 25: ...lacement unscrew and remove the cover of the battery compartment remove the old battery insert the new battery observing the polarity the sign of the battery should be inwardly directed fix the cover...

Страница 26: ...Dosimeter does not respond to pressing a button the LCD indicates incorrect symbols Microprocessor error condition Remove the battery and insert it again in 5 minutes 4 The LCD indicates Er1 Er7 Dosi...

Страница 27: ...no more than 5 at confidence probability 0 95 5 5 Barometer Scale interval 1 kPa Measurement range 60 120 kPa 5 5 Thermometer Scale interval 0 1 Measurement range 10 30 5 5 Hydrometer Measuring range...

Страница 28: ...nt operation Instrument is considered as not valid for further verification if it doesn t meet the above requirements 5 7 2 Carrying out verification testing Testing requires following check instrumen...

Страница 29: ...0 s interval at least Then calculate average DER value j H by formula 5 1 5 1 i ji j H H 5 2 where ji H i th dosimeter reading when measuring DER in j th tested point 6 repeat measurements for points...

Страница 30: ...urement main relative measurement error is calculated as follows 1 set maximum DER and DE threshold levels on the dosimeter and switch on DE measurement mode 2 do actions described in 5 7 3 1 2 3 read...

Страница 31: ...lastic bags with silica gel 6 4 When carried by air Dosimeters in package should be placed in hermetic compartments 6 5 Dosimeters should be stored in the manufacturer s package at the air temperature...

Страница 32: ...of the Quality Control Department of the manufacturer 7 4 Warranty and after warranty repair is carried out by the manufacturer or the institutions that have a permission of the manufacturer 7 5 Warra...

Страница 33: ...bration direction 0o Figure A 1 for horizontal installation at the phantom rotation in vertical plane 1 241 Am 59 5 keV 2 137 Cs 662 keV 3 60 Co 1250 keV Figure A 2 for vertical installation at the ph...

Страница 34: ...ATTACHMENT B reference DIAGRAM OF DOSIMETER ROTATION TO MEASURE ANGULAR RESPONSE Figure B 1 Diagram of dosimeter rotation in horizontal plane Figure B 2 Diagram of dosimeter rotation in vertical plan...

Страница 35: ...35 ATTACHMENT C reference OVERALL DIMENSIONS EFFECTIVE CENTER OF THE DOSIMETER DETECTOR Figure C 1...

Отзывы: