Calibration 

TK100C

The output voltage of the TK100C can be adjusted. 

On the bottom side of the device is a trimmer potentiometer located under the identification label.

It is marked by the "DC Adjust" symbol.

Please note, that any previous calibration is invalidated by adjustments you make to this trimmer.

Safety Precautions & Override 

   

 

The TK100C has a safety circuit inside, wich enables output voltage only if a BNC Adapter is connected.

You can override this safety feature at your own risk following these steps:

1. Disconnect the TK100C from the 15 V DC supply voltage.

2. Wait for at least 20 seconds.

3. Hold down the 

Mode Button

 (5) while reconnecting the 15 V DC supply voltage.

The safety circuit is now circumvented.

You can reenable the safety circuit by repeating the steps above, once again. 

Specifications 

Squarewave output voltage : 

+100 V ± 0.5%

Risetime : 

< 6 ns

Overshoot : 

< 3%

Repetition Rate : 

100 Hz

Pulse Width: 

5 ms

Dimensions: 

135 x 30 x 70 mm

Weight: 

120 g

Environmental Specifications

Altitude   

 

operating   

up to 2000 m

 

 

 

non-operating 

up to 15000 m

Temperature Range   

operating   

0° C to +50° C

 

 

 

non-operating 

-40° C to +71° C

Maximum Relative Humidity 

operating   

80 % relative humidity for temperatures up to +31° C, 

 

 

 

 

 

decreasing linearly to 40 % at +50° C

Содержание TK100C

Страница 1: ...TK100C Probe Calibrator Instruction Manual...

Страница 2: ...om date of shipment and will repair or replace any defective product which was not dama ged by negligence misuse improper installation accident or unauthorized repair or modification by the buyer This...

Страница 3: ...ic product at the end of its life either to the manufacturer or take care of separate WEEE collection and professional WEEE treatment yourself Do not dispose as unsorted municipal waste EC Directives...

Страница 4: ...are present when using this instrument Use extreme caution when handling testing and adjusting Use only grounded instruments Connect the generator only to equipment that is connected to earth ground A...

Страница 5: ...tabilized 100 V DC source loads a 1 uF capacitor A none chatter contact connects the voltage to a RC network of approx 50 and 47 pF The rise time between 10 and 90 is approx smaller 6 ns The duration...

Страница 6: ...e TK100C provides a 100 V square wave signal of 5 ms duration NOTE The TK100C is in pulse mode The Mode Indicator LED 3 blinks By pressing the Mode Button 5 again the TK100C is set to DC Mode providin...

Страница 7: ...Disconnect the TK100C from the 15 V DC supply voltage 2 Wait for at least 20 seconds 3 Hold down the Mode Button 5 while reconnecting the 15 V DC supply voltage The safety circuit is now circumvented...

Страница 8: ...y Please check the delivery for completeness If any item is missing send a message to our sales department and we will send you this item immediately Item Qty BNC Adapter 2 5 1 BNC Adapter 5 0 1 BNC A...

Отзывы: