PMK TK100C Скачать руководство пользователя страница 6

6

How to use this instrument 

TK100C

Setup:

1. Connect the probe calibrator to the power supply and the power supply to mains.

  NOTE: The 

Power LED

 (2) is lit, 

  The 

Mode Indicator

 

LED

 (3) is off. 

  There is no voltage applied to the 

BNC

 

Output

 (4)

2. Connect the probe to the  BNC jack of the measuring instrument and adjust the scope-to-probe 

  divider ratio.

4. Connect the probe to the 

BNC

 

Output 

(4) of the TK100C

Operation:

- By pressing the 

Mode Button

 (5) once, the TK100C provides a 100 V  square wave signal of 5 ms duration.

  NOTE: The TK100C is in pulse mode. The 

Mode Indicator LED

 (3) blinks.

- By pressing the 

Mode Button

 (5) again the TK100C is set to 

DC Mode,

 providing 100 V DC

  NOTE: The 

Mode

 

Indicator LED 

(3) glows constantly.

- If the 

Mode Button

 (5) is pressed again the TK100C will go into 

Standby

 once again.

  NOTE: The 

Mode Indicator LED

 (3) is turned off again.

Make sure to set the trigger level at the scope to a decent value for best results. 

( e.g. 50% of the pulse amplitude) 

1  Input Supply Voltage 15 V DC
2  Power LED
3   Mode LED
4  Output BNC Connector

5  Mode Button
6  Trimmer Potentiometer

7  Identification Label

Содержание TK100C

Страница 1: ...TK100C Probe Calibrator Instruction Manual...

Страница 2: ...om date of shipment and will repair or replace any defective product which was not dama ged by negligence misuse improper installation accident or unauthorized repair or modification by the buyer This...

Страница 3: ...ic product at the end of its life either to the manufacturer or take care of separate WEEE collection and professional WEEE treatment yourself Do not dispose as unsorted municipal waste EC Directives...

Страница 4: ...are present when using this instrument Use extreme caution when handling testing and adjusting Use only grounded instruments Connect the generator only to equipment that is connected to earth ground A...

Страница 5: ...tabilized 100 V DC source loads a 1 uF capacitor A none chatter contact connects the voltage to a RC network of approx 50 and 47 pF The rise time between 10 and 90 is approx smaller 6 ns The duration...

Страница 6: ...e TK100C provides a 100 V square wave signal of 5 ms duration NOTE The TK100C is in pulse mode The Mode Indicator LED 3 blinks By pressing the Mode Button 5 again the TK100C is set to DC Mode providin...

Страница 7: ...Disconnect the TK100C from the 15 V DC supply voltage 2 Wait for at least 20 seconds 3 Hold down the Mode Button 5 while reconnecting the 15 V DC supply voltage The safety circuit is now circumvented...

Страница 8: ...y Please check the delivery for completeness If any item is missing send a message to our sales department and we will send you this item immediately Item Qty BNC Adapter 2 5 1 BNC Adapter 5 0 1 BNC A...