11
reduction and metal rust of electrical components. Be sure not to clean the medical refrigerator with
hot water and corrosive detergent or organic solvent. Also do not clean the medical refrigerator with
scrubbing brush and wire brush. Children are not allowed to play games with the medical refrigerator.
Notes:Do not allow children to use medical freezer to playing games .
Notes:Not to put heavy objects on the refrigerator door or medical cap, to avoid the door or the
top compression deformation.
V. Clearing of Fault and Maintenance Service
Some abnormal conditions of the medical refrigerator are caused by misuse. Before you commit maintenance,
please compare with the following table for self inspection and exclusion.
Problems
Reasons and Solving Measures
Medical refrigerator does not work
·If the power socket has electricity?
·If the power socket is plugged or loosened?
·If the power fuse is disconnected?
·If the supply voltage is too low or too high?
Compressor breaks down
·If the temperature setting is right?
Temperature goes on decreasing
after reaching to the set value
·If the temperature setting is right?
Temperature cannot reach the set
value
·If the fan stops running?
·If the door is not closed tight or opened too frequently?
·If too many goods are put in at one time and if the air channel is
blocked off?
·If the environment temperature is too high?
Too much noise
·If the refrigerator body is placed at the flat ground?
·If the refrigerator body touches the wall?
·If the incubator immediately enters operating state after being
started?
If you cannot judge the fault cause or debug, please contact the local or nearby appointed service center
according to the instruction of the warranty certificate (standard accessory) and tell use your name, address,
phone number, type of the medical refrigerator, purchase time, fault phenomenon, etc. Professional engineer