S = Small adult
•
M = Medium-sized adult
•
L = Large adult
•
XL = Extra large adult
The preset exposure values are shown below the patient sizes.
NOTE
The exposure values will automatically change according to the selected
patient size, image resolution and ULD setting. The preset exposure
values are shown in section "Adjusting exposure values for current
11.6 Adjusting image volume position and size
About this task
Follow these instructions to adjust image volume position and size.
NOTE
This function is not available for the Skull program.
You see two camera images of the patient’s head on the control panel: a
front view and a side view. The preset position and size of the image volume
are shown with a blue area in both views. You can use either or both views
to adjust the volume position and size.
NOTE
The blue area is not an exact representation of the image volume. It is
intended for visualization purposes only.
Steps
1. To adjust the volume position, use your mouse cursor (virtual control
panel) or finger (touch screen) to move the blue area to the anatomical
region that you wish to expose.
The blue area can be moved in any direction.
2. Adjust the volume size on the touch screen smaller or larger by using
pinch gestures (1), or holding your mouse cursor (virtual control panel) or
11 3D exposure
58 Planmeca Viso
User's manual
Содержание Viso
Страница 1: ...Planmeca Viso user s manual EN 30005931 30005931...
Страница 10: ...Warning hot surface Standard ISO 7010 General warning Standard ISO 7010 4 Symbols 6 Planmeca Viso User s manual...
Страница 45: ...Use the segmentation buttons to select the active exposure areas 10 2D exposure User s manual Planmeca Viso 41...
Страница 79: ...Results The image is shown on the computer screen 11 3D exposure User s manual Planmeca Viso 75...
Страница 102: ...Patient supports 16 Cleaning and disinfection 98 Planmeca Viso User s manual...
Страница 110: ......
Страница 111: ......