11.7 Selecting image resolution, Ultra Low Dose (ULD) and ProFace
Steps
1. Select field (1) to open a pop-up window.
2. Select the image resolution you wish to use (2). The voxel size options
are 600 / 450, 300, 200, 150 and 75 micrometres, depending on the
selected program.
3. Select the ULD (Ultra Low Dose) button (3) if you want to take an
exposure with a very low dose. The voxel size buttons turn green and
show a small apple.
4. Select the ProFace (optional feature) button (4) to add the 3D photo to
the exposure.
5. Select the green check mark button (5).
NOTE
The available options depend on the selected program.
NOTE
The exposure values will automatically change according to the selected
patient size, image resolution and ULD setting. The preset exposure
values are shown in section "Adjusting exposure values for current
11.8 Adjusting exposure values for current exposure
About this task
NOTE
Always try to minimise the radiation dose to the patient.
The exposure values have been preset at the factory for each patient size,
image resolution and ULD (Ultra Low Dose) setting. The preset exposure
values are average values and they are only meant to guide the user.
The preset exposure values are shown in the following tables.
11 3D exposure
60 Planmeca Viso
User's manual
Содержание Viso
Страница 1: ...Planmeca Viso user s manual EN 30005931 30005931...
Страница 10: ...Warning hot surface Standard ISO 7010 General warning Standard ISO 7010 4 Symbols 6 Planmeca Viso User s manual...
Страница 45: ...Use the segmentation buttons to select the active exposure areas 10 2D exposure User s manual Planmeca Viso 41...
Страница 79: ...Results The image is shown on the computer screen 11 3D exposure User s manual Planmeca Viso 75...
Страница 102: ...Patient supports 16 Cleaning and disinfection 98 Planmeca Viso User s manual...
Страница 110: ......
Страница 111: ......