Planmeca SingLED operating light 13
TECHNICAL SPECIFICATIONS
User’s Manual
Conformance with standard ISO 9680-2007
The measured sample does not fulfil the following parts of ISO/FDIS 9680:2007(E)
standard:
5.2.7 Shadow: the hard shadow is too large in vertical and horizontal direction.
5.2.8 Colour rendering index: Ra is too low for diagnostic purposes.
5.2.9 UV irradiance: not measured 180nm to 200nm, because limition of measuring
device.
Содержание SingLED
Страница 1: ...user s manual 10020003_8 EN...
Страница 2: ......
Страница 18: ...TECHNICAL SPECIFICATIONS 14 Planmeca SingLED operating light User s Manual Dimensions 2 4 2 4 115 248 166...
Страница 19: ...Planmeca SingLED operating light 15 TECHNICAL SPECIFICATIONS User s Manual 1 1 4 9 0 9 0 0 6 3 217 174...
Страница 23: ...Planmeca SingLED operating light 19 TECHNICAL SPECIFICATIONS User s Manual Dimensions of the wall plate in mm...
Страница 26: ......