INSTALLATION OF THE PLANMECA ETHERNET INTERFACE BOX TO THE DIXI3 DIGITAL INTRA X-RAY SYS-
4 Planmeca Ethernet interface
Installation manual
Installing the Intra sync cable
The sync cable synchronizes the exposure and receiving the image, i.e. detects the start of
the exposure. The sync cable is required when you want to store the exposure parameters
with the image.
It is also required when you want to use the Planmeca Intra control panel in the
Dimaxis 4.0
software
Intraoral Exposure
window. The exposure parameters can be set in this window if
you are using the Planmeca Intra with updated software 3.0.3. The control panel logic is iden-
tical to that of the Planmeca Intra remote control.
If you are NOT connecting any other equipment to the Ethernet interface, use the Planet
grabber cable with one connector. In this case connect the interconnection cable from Dixi3 to
Port
1
.
If you are using the Planet grabber cable with two connectors, note that the connectors are
numbered. If you connected the interconnection cable to port
1
connector, you must connect
the 10m/male/female cable to the Planet connector number
1
, and if you connected the inter-
connection cable to port
2
connector, you must connect the 10m/male/female cable to the
Planet connector number
2
.
Connect the 10m/male/female cable between the Planet connector on the generator box and
Planet grabber cable.
Installing the Dimaxis software
Install the Dimaxis Pro or Dimaxis Classic software to your computer in case it is not yet
installed (refer to Dimaxis Pro Installation manual, publication number 10006119, or Dimaxis
Classic Installation manual, publication number 10006182).
Planet grabber cable and
sync cable