Chapter D - HELP & ERROR MESSAGES
Planmeca Compact i D-19
DETAILED DESCRIPTION OF ERROR MESSAGES
Technical Manual
3.8 Operating light related error messages
E 56
OP-light erroneously connected or short circuit in push-
button
This error can occur only during selftest at power up. If the momentary push button is shorted
during the test then this error is generated. The button could be damaged, or the reason can
be a wrong connection so that the bulb and button are cross-connected. The electronics
refuses to turn on the output power to the operating light, until the short has been removed (or
the connection has been corrected). This feature protects the OP-light push button from
destruction. Obviously the button should not be pressed simultaneously when powering on
the unit, to avoid a false error.
E 57
reserved
E 58
Operating light bulb burned out
This error occurs when the operating light bulb has burned out. Replace the bulb.
3.9 Separator system related
E 59.n The separator is reporting an functional error
The separator is signalling the error output (either Metasys (59.1) or Dürr (59.2)). Please
check the separator documentation.
E 60
The separator is not responding or the separator cable is
faulty
The stand-by signal is missing from the separator (valid only for Dürr or Metasys). Either the
separator is faulty, or the cable between the MCB and separator is faulty. Check that the sep-
arator is reaching a power supply voltage. Check that the separator cable is plugged in to the
MCB properly.
E 61
Separator is flooded with too much water
This error can occur only if the unit is equipped with a Microvac separator. The error is gener-
ated if the water level inside the Microvac rises too high. The reason can be a too fast water
flow to the separator compared with the handling capacity of the drain. Check the drain, clean
the bottom cup under the Microvac. If necessary, replace the Microvac separator.
E 62
The separator is unknown
The newly attached separator is of a new type and cannot be controlled by the MCB. Please
update the MCB software. Check that the separator connector is properly connected and
none of the wires are loose.
E 63
WMS error
WMS error:
E 63.1
There is WMS installed but there should not be (according to service mode)
E 63.2
There is no WMS installed but there should be according to service mode
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......