Planmeca Compact i F-1
Chapter
Technical Manual
F
ADJUSTMENTS
1 MECHANICAL ADJUSTMENTS
1.1 Adjusting vertical position of the OP delivery arm column
The console arm column must be vertical. You may have to adjust its position, if the floor is
very inclined.
Adjustment screws
Console arm adapter
To adjust the vertical position of the console arm col-
umn:
1. Remove the bowl.
2. Remove the cuspidor covers.
3. Adjust the position of the console arm adapter with
the three M8x12 DIN 913 adjustment screws sup-
plied with the mounting accessories. Refer to the fig-
ures for the location of the three holes.
4. Reassemble in reverse order.
Console arm
column
Console arm
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......