Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
FC.LD Move foot control pedal to leftmost position.
Press pedal down and hold. While holding,
push centre knob briefly in 'chair up'
direction.
Move the foot control pedal to the leftmost
position. Press the pedal down and hold.
While holding, push the centre knob briefly
in the ‘chair up’ direction.
FC.LU Move foot control pedal to leftmost position
and hold. While holding, push centre knob
briefly in 'chair up' direction.
Move the foot control pedal to the leftmost
position and hold. While holding, push the
centre knob briefly in the ‘chair up’ direction.
FC.CD Press foot control pedal down and hold.
While holding, push centre knob briefly in
'chair up' direction.
Press the foot control pedal down and hold.
While holding, push the centre knob briefly
in the ‘chair up’ direction.
FC.CU Keep foot control pedal in rest position
(centre). Push centre knob briefly in 'chair
up' direction.
Keep the foot control pedal in the rest
position (centre). Push the centre knob
briefly in the ‘chair up’ direction.
FC.RD Move foot control pedal to rightmost
position. Press pedal down and hold. While
holding, push centre knob briefly in 'chair up'
direction.
Move the foot control pedal to the rightmost
position. Press the pedal down and hold.
While holding, push the centre knob briefly
in the ‘chair up’ direction.
FC.RU Move foot control pedal to rightmost position
and hold. While holding, push centre knob
briefly in 'chair up' direction.
Move the foot control pedal to the rightmost
position and hold. While holding, push the
centre knob briefly in the ‘chair up’ direction.
FC.--
To exit foot control calibration mode, press
and hold down calibration switch for 4
seconds.
H 7003 Control panel is temporarily disabled due to
interference on touch screen.
Make sure the touch screen is clean and
dry. If the problem continues, contact
service.
31.3 Error messages in short form
The error message tables list the error messages, which are mainly intended
to assist the technician. If an error message is displayed, contact service.
For more detailed information, see
Planmeca Compact i Touch technical
manual.
31.3.1 General error messages
General error messages
ERROR
CODE
EXPLANATION
E1.01
Main water valve short circuit.
Disconnect valve and check if error disappears.
E1.02
Main air valve short circuit.
Disconnect valve and check if error disappears.
E1.03
Bowl rinse valve short circuit.
Disconnect valve and check if error disappears.
31 Help and error messages
264 Planmeca Compact i
User's manual
Содержание Compact i Touch v2
Страница 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Страница 10: ...Table of contents Planmeca Compact i User s manual ...
Страница 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Страница 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Страница 307: ...34 3 5 Side delivery with hanging tube instruments 34 Technical information User s manual Planmeca Compact i 297 ...
Страница 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Страница 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Страница 313: ......