6-4
Chapter 6: Maintaining The Integra
™
Model 9K0U Scale
Integra
™
Platform Scale 9K0U Operator Guide SV61955 Rev. A
4. Press
Enter
and the Service Setup menu appears.
5. Press the down cursor key to advance to the next page of
the Service Setup menu, and select Metrology Setup. The
Metrology Setup menu appears.
6. Select
Scale Calibration
and follow the prompts displayed
on the scale. When you are done, the Metrology Setup
menu reappears in the display.
7. Select
Exit
,
Exit
,
Exit
, then select
Save Changes
. The
scale resets and is now calibrated. If you place the 20
lb. weight on the scale, “19 lb. 15.8 oz.” or “20 lb. 0.0 oz”
should appear in the display.
Calibrating the
9K0U Scale
(20 lb. Calibration
Weight Required)
Содержание 9K0U
Страница 1: ...Operator Guide US English Version Integra Platform Scale 9K0U...
Страница 12: ...This Page Intentionally Blank...
Страница 18: ...This Page Intentionally Blank...
Страница 34: ...This Page Intentionally Blank...
Страница 38: ...This Page Intentionally Blank...
Страница 47: ...Integra Model 9K0U Scale Operator Guide SV61955 Rev A Integra Platform Scale 9K0U Operator Guide Index...