PIE piecal 334 Скачать руководство пользователя страница 6

Accessories

AC ADAPTOR (200 to 240 VAC)

 

 

 

Part No. 020-0100

AC ADAPTOR (100 to 120 VAC)

 

 

 

Part No. 020-0101

Ni-MH 1 Hour Charger w/4 Ni-MH AA Batteries

 

Part No. 020-0103

Guarantee

The Practical Instrument Electronics PIECAL Model 334 is guaranteed to be a functional replacement 

for the Altek Model 334 or Altek Model 334A as described in the product comparison. Claims under 

this guarantee can be made by returning the equipment prepaid to our factory. The equipment will 

be  repaired,  replaced,  adjusted  or  money  back  at  our  option. The  liability  of  Practical  Instrument 

Electronics (PIE) is restricted to that given under our guarantee. No responsibility is accepted for 

damage, loss or other expense incurred through sale or use of our equipment. Under no condition 

shall  Practical  Instrument  Electronics,  Inc.  be  liable  for  any  special,  incidental  or  consequential 

damage.

Warranty

Our equipment is warranted against defective material and workmanship (excluding batteries) for a 

period of three years from the date of shipment. Claims under warrantee can be made by returning 

the equipment prepaid to our factory. The equipment will be repaired, replaced or adjusted at our 

option. The  liability  of  Practical  Instrument  Electronics  (PIE)  is  restricted  to  that  given  under  our 

warrantee. No responsibility is accepted for damage, loss or other expense incurred through sale or 

use of our equipment. Under no condition shall Practical Instrument Electronics, Inc. be liable for any 

special, incidental or consequential damage.

SETTING UP VALVES

When setting up a valve it is important to correctly set the end stops. Use the PIECAL Model 334 wo 

supply the 4 to 20 mA control signal to stroke the valve. Select “SOURCE” and the PIECAL Model 

334 will use the internal power source fpr outputting current or switch to 2-WIRE SIMULATOR to 

stroke a valve using any pre-existing installed loop power supply as the power source.

Example:

1)  Disconnect the 4-20 mA control wires from the Current-to Pressure (I/P) converter or the actuator.

2)  Connect  the  PIECAL  Model  334  following  the  connection  diagrams  on  the  prevous  pages  for 

Simulate 2-Wire Transmitters

3)  Move the EZ-CHECK™ switch 

e

 to  “4.00 mA”/”0.0%”

 and adjust the fully closed stop on the 

actuator.

4)  Turn the PIECAL Model 334’s knob 

slowly counterclockwise and verify that the actuator and 

valve don”t move. Repeat steps 3 & 4 until no movement is detected.

5)  Move the EZ-CHECK™ switch 

to DIAL  and quickly back to  “4.00 mA”/”0.0%”

 then turn the 

PIECAL Model 334’s knob 

clockwise. The actuator and valve should begin to move.

6) Move the EZ-CHECK™ switch 

to “20.00 mA”/”100.0%” 

and adjust the fully open stop on the 

actuator.

7)  Turn the PIECAL Model 334’s knob 

slowly clockwise and verify that the actuator and valve 

don”t move. Repeat steps 6 & 7 until no movement is detected.

8)  Move the EZ-CHECK™ switch 

to DIAL  and quickly back to  “20.00 mA”/”100.0%”

 then turn 

the PIECAL Model 334’s knob counterclockwise. The actuator and valve should begin to move

.

Application Notes

Practical Instrument Electronics, Inc.

Available at www.Instrumentation.com 

Содержание piecal 334

Страница 1: ...to check loop response and control settings The PIECAL Model 334 uses any loop power from 2 to 100V DC READ LOOP CURRENT Check controller outputs or measure the milliamp signal anywhere in the loop T...

Страница 2: ...displayed in the units selected r DIAL KNOB Turn the knob to adjust output level Turn clockwise to increase the output counter clockwise to decrease the output t EXTERNAL POWER JACK Not Shown When us...

Страница 3: ...while the EZ CHECK switch e is in the DIAL position or the current can be set at the xed points of 4 00mA 0 0 or 20 00mA 100 0 with switch e READ mA READ Percent of 4 to 20 mA Choose this function to...

Страница 4: ...IECAL Model 334 to the plus input of the device and the black source lead to the minus The PIECAL Model 334 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the...

Страница 5: ...RCE output to full scale supplies a nominal 24V DC to power a 2 Wire Transmitter while simultaneously displaying the 4 to 20 mA output of the transmitter READ MILLIAMPS Select READ milliamps by moving...

Страница 6: ...Electronics Inc be liable for any special incidental or consequential damage SETTING UP VALVES When setting up a valve it is important to correctly set the end stops Use the PIECAL Model 334 wo supply...

Страница 7: ...dication with nominal 1 hour of operation left Protection against misconnection Over voltage protection to 135 vrms rated for 30 seconds or 240 vrms rated for 15 seconds Display High contrast backlit...

Страница 8: ...V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 100 VDC fuse less protected from reverse polarity connections Battery life 125 hours nominal Voltage Read Range and...

Отзывы: