Product Description

•  

EASY TO USE

With the PIECAL Model 334 you can check, calibrate 

and measure all your current signal instruments in a 

4 to 20 milliamp DC loop. It can be used at any 

access point in your loop. Source & Read 0.00 to 

24.00 mA, Simulate a 2 Wire Transmitter or use the 

PIECAL Model 334 to simultaneously power your 2 

Wire Transmitter and measure its output. When 

desired the PIECAL Model 334 can display current 

in milliamps or percent of 4 to 20.

•   SOURCE MILLIAMPS

Calibrate recorders, digital indicators, stroke valves 

or any instruments that get their input from a 4 to 

20 mA loop. Easily set any value quickly to within 

0.01 mA with the adjustable digital potentiometer 

DIAL ” or use preset  4.00 mA (0.0%)  and  20.00 

mA (100.0%)  EZ-CHECK™ settings.

•   RECALL OUTPUT SETTINGS

The EZ CHECK™ switch provides rapid checking of 

4.00, 20.00 and any convenient third point between 

0.00 to 24.00 mA.

•  CALIBRATE USING LOOP POWER

Check loop wiring and receivers by using the 

PIECAL Model 334 in place of a 2 Wire transmitter. 

Simulate a changing process input to check loop 

response and control settings. The PIECAL Model 

334 uses any loop power from 2 to 100V DC.

•   READ LOOP CURRENT

Check controller outputs or measure the milliamp signal anywhere in the loop. The 

PIECAL Model 334 measures 0.00 to 52.00 mA signals with greater accuracy than 

a typical multimeter. The PIECAL Model 334 can be easily switched to display 

milliamps or percent of 4 to 20.

•   POWER & MEASURE 2 WIRE TRANSMITTERS

The PIECAL Model 334 can simultaneously output 24V DC to power any and all devices in a 

process loop using the internal batteries and the internal switching power supply, while 

measuring the output of a 2 Wire Transmitter and any other loop devices. This is handy for 

checking the functionality of transmitters in the  eld or on the bench.

•   READ DC VOLTS

The PIECAL Model 334 can measure from -99.99 to +99.99 VDC with 10mV resolution. Use it to 

check loop power supplies, I/V converters, chart recorders, 1 to 5 Volt signals, and any other 

voltages within this range making it unnecessary to carry an additional multimeter.

PIECAL Model 334

4-20 mA Loop Calibrator

Operating Instructions

PIECAL Model 334

4-20 mA Loop Calibrator

Operating Instructions

Practical Instrument Electronics, Inc.

Available at www.Instrumentation.com 

Содержание piecal 334

Страница 1: ...to check loop response and control settings The PIECAL Model 334 uses any loop power from 2 to 100V DC READ LOOP CURRENT Check controller outputs or measure the milliamp signal anywhere in the loop T...

Страница 2: ...displayed in the units selected r DIAL KNOB Turn the knob to adjust output level Turn clockwise to increase the output counter clockwise to decrease the output t EXTERNAL POWER JACK Not Shown When us...

Страница 3: ...while the EZ CHECK switch e is in the DIAL position or the current can be set at the xed points of 4 00mA 0 0 or 20 00mA 100 0 with switch e READ mA READ Percent of 4 to 20 mA Choose this function to...

Страница 4: ...IECAL Model 334 to the plus input of the device and the black source lead to the minus The PIECAL Model 334 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the...

Страница 5: ...RCE output to full scale supplies a nominal 24V DC to power a 2 Wire Transmitter while simultaneously displaying the 4 to 20 mA output of the transmitter READ MILLIAMPS Select READ milliamps by moving...

Страница 6: ...Electronics Inc be liable for any special incidental or consequential damage SETTING UP VALVES When setting up a valve it is important to correctly set the end stops Use the PIECAL Model 334 wo supply...

Страница 7: ...dication with nominal 1 hour of operation left Protection against misconnection Over voltage protection to 135 vrms rated for 30 seconds or 240 vrms rated for 15 seconds Display High contrast backlit...

Страница 8: ...V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 100 VDC fuse less protected from reverse polarity connections Battery life 125 hours nominal Voltage Read Range and...

Отзывы: