![PIE 820-ELITE Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/pie/820-elite/820-elite_operating-instructions-manual_1555580008.webp)
Page 6
SOURCE
HI
SET
LO
OFF
MAX
READ
MIN
READ
OVERLOAD
Hz SYNC
Double Click
Menu
Push & Turn
for
Fast Dialing
Push & Hold
to
Store/Step
mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure
READ
HI
TYPE K °C
820.1
q
EZ-CHECK™ SWITCH
SOURCE:
Instantly output two preset settings by moving the
EZ-CHECK™ switch to the
“LO”
position or
“HI”
position. For fast
three point checks select the
“SET”
position. The PIE 820-ELITE
will remember the last
“SET”
value, even with the power off.
These values can easily be changed to suit the calibration requirements.
The values stored in the HI and LO positions are also used for Auto
Stepping.
READ:
Slide the switch to the SET position. The PIE 820-ELITE will
display the current reading from the sensor or device being measured.
Slide the switch to MAX and the highest value measured since turn-on
or reset will be displayed; slide the switch to MIN and the lowest value
measured since turn-on or reset will be displayed.
Operating Instructions
Basic Operation
www.
GlobalTestSupply
.com
Quality PIECAL Products Online at: