READ

HI

2 WIRE SIM  mA

20.000

SOURCE

HI

SET

LO

OFF

MAX
READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click 

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure

To

Sensor

Typical

2-Wire

Transmitter

(Disconnected)

+ IN -

REF +OUT-

Receiver

Powers External

2-Wire Transmitter 

Power Supply

(2 to 100 VDC) 

2 Wire SIM mA, 2 Wire SIM % (Percent of 4 to 20 mA)

Choose this function to simulate a 2 Wire Transmitter output from 
0.000 to 24.000 milliamps. Operates in loops with power supply 
voltages from 2 to 60 VDC.

Move the power switch 

to SOURCE then Double Click the 

EZ-DIAL knob to get into the Double Click Menu. Turn the knob 
to scroll through the settings and press the knob to make your 
selection. Select mA for the FUNCTION and 2W SIM for the 
MODE. Choose either mA or % and whether you need the 250

 

HART resistor active in the loop.

Connect the output leads of the PIE 820-ELITE to the inputs of 
the device being calibrated, making sure to check polarity. Red 
lead to the plus (+) input and black lead to the minus (-) input.

Instantly output your SPAN and ZERO output settings by moving 
the EZ-CHECK switch between HI and LO (defaults to 20 & 4 
mA). You may also select any third output setting (such as 
mid-range) using the SET position on the EZ-CHECK switch. 
The output is adjusted in 0.001 mA (0.01%) increments by 
turning the knob 

e

. Press and turn the knob for faster dialing 

with 0.100 mA (1.00%) increments.

www.

GlobalTestSupply

.com

Quality PIECAL Products Online at:

[email protected]

Содержание 820-ELITE

Страница 1: ...ifunction Process Calibrator mA V pH TC RTD Freq Pressure Operating Instructions Practical Instrument Electronics 82 East Main Street Suite 3 14 Webster NY 14580 USA Tel 585 872 9350 Fax 585 872 2638...

Страница 2: ...easure Transmitters with Leak Detection 16 Voltage Millivolt Source V mV 18 Read V mV 19 pH Source pH 20 Thermocouple Source T C 22 Read T C Sensors 23 Resistance Source Resistance 24 Read Resistance...

Страница 3: ...99 Hz 0 0 to 9999 9 Hz and 0 000 to 20 000 kHz Checkout resistance instruments loop wiring problems Source and read resistance from 0 00 to 410 00 and 0 0 to 4001 0 ohms Beep out connections with the...

Страница 4: ...gs for instant recall with the EZ CHECK switch Choose between 2 3 5 11 steps or RAMP to increase and decrease the output between Zero and Span from 5 6 7 8 9 10 15 20 25 30 and 60 seconds Measure temp...

Страница 5: ...ck Leads with Banana Plugs Spade Lugs OPTIONAL Ni MH 1 Hour Charger with 4 Ni MH AA Part No 020 0103 Batteries 100 120 V AC input for North America Only T C Wire Kit 1 for Types J K T E Part No 020 02...

Страница 6: ...ill automatically turn off To change the batteries remove the rubber boot remove the battery door from the back of the unit by sliding the door downward This allows access to the battery compartment R...

Страница 7: ...t value is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 3 Instantly output your SPAN and ZERO output settings by moving the EZ CHECK...

Страница 8: ...with the power off These values can easily be changed to suit the calibration requirements The values stored in the HI and LO positions are also used for Auto Stepping READ Slide the switch to the SET...

Страница 9: ...ut value changes SELECTING FUNCTIONS The EZ DIAL knob is used to setup the PIE 820 ELITE to match the instrument to be calibrated or signal to be measured Each time you turn on the PIE 820 ELITE the L...

Страница 10: ...s the Double Click Menus to select each function and the options for each function Available choices are shown in grey EXIT FUNCTION V RANGE 10V 1V 60V 100mV Source mA Simulate 2 Wire Transmitters Rea...

Страница 11: ...0 a 4274 Cu 50 a 4280 Ni 120 a 6720 Ni 110 a 5801 Source Read Thermocouples Source RTD Operating Instructions Double Click Menus Read RTD EXIT FUNCTION OHMS RANGE 400 4000 WIRE MODE 234W EXIT FUNCTION...

Страница 12: ...ings are remembered even with the power off EXIT MENU exits this menu immediately and saves any changes Menu will automatically exit after 15 seconds of inactivity AUTO OFF If AUTO OFF is ON the unit...

Страница 13: ...using the calibrator in areas with enough ambient light to read the display STEPS pressing the knob will cycle through 2 3 5 11 and RAMP The endpoints of the steps or ramp are based on the values stor...

Страница 14: ...settings and press the knob to make your selection Select mA for the FUNCTION and SOURCE for the MODE Choose either mA or and whether you need the 250 HART resistor active in the loop Connect the outp...

Страница 15: ...croll through the settings and press the knob to make your selection Select mA for the FUNCTION and 2W SIM for the MODE Choose either mA or and whether you need the 250 HART resistor active in the loo...

Страница 16: ...either mA or and whether you need the 250 HART resistor active in the loop Connect the red input lead of the PIE 820 ELITE to the more positive point of the break and the black input to the more negat...

Страница 17: ...he settings and press the knob to make your selection Select mA for the FUNCTION and PWR MEAS for the MODE Choose either mA or and whether you need the 250 HART resistor active in the loop Disconnect...

Страница 18: ...the 820 ELITE to the plus input of the device and the black source lead from the mA to the minus The PIE 820 ELITE supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The curren...

Страница 19: ...er and is leaking somewhere in the loop LEAK ERROR OVER RANGE PWR MEAS LEAK ERROR 00 28 12 280mA PWR MEAS Typical Error Conditions READ SOURCE HI SET LO OFF MAX READ MIN READ OVERLOAD Hz SYNC Double C...

Страница 20: ...Turn the knob to scroll through the settings and press the knob to make your selection Select V for the FUNCTION and 10V or 80 mV for the RANGE Connect the output leads of the PIE 820 ELITE to the in...

Страница 21: ...ress the knob to make your selection Select V for the FUNCTION and 10V 1V 60V or 100 mV for the RANGE Connect the red input lead of the PIE 820 ELITE to the more positive point of the break and the bl...

Страница 22: ...black lead from the mV jack to the minus input You may also use the optional adaptor jack 020 0 if you require BNC connections Instantly output your SPAN and ZERO output settings by moving the EZ CHEC...

Страница 23: ...rating correctly The 820 ELITE simulates 0 000 to 14 000 pH 25 C corresponding to 414 12 to 414 12 mV Once the pH instrument has been adjusted against the PIE 820 ELITE reconnect the pH probe and chec...

Страница 24: ...the proper type of thermocouple wire via the miniature thermocouple socket Instantly output your SPAN and ZERO output settings by moving the EZ CHECK switch between HI and LO You may also select any...

Страница 25: ...e socket If no sensor is connected a wire is broken or the sensor is burned out OPEN TC will appear on the display Signals above the maximum scale are limited by protection circuitry with OVERRANGE on...

Страница 26: ...wires Instantly output your SPAN and ZERO output settings by moving the EZ CHECK switch between HI and LO You may also select any third output setting such as mid range using the SET position on the...

Страница 27: ...4 wires If continuity is selected resistance is measured up to 400 0 The beeper will sound and appears on the display when resistances below 100 0 are measured Signals above the maximum scale are lim...

Страница 28: ...onnect the PIE 820 ELITE to the inputs of the device using 2 3 or 4 wires Instantly output your SPAN and ZERO output settings by moving the EZ CHECK switch between HI and LO You may also select any th...

Страница 29: ...all wires from the sensor and connect the PIE 820 ELITE to the inputs of the device using 2 3 or 4 wires Signals above the maximum scale are limited by protection circuitry with OVERRANGE flashed on...

Страница 30: ...pickups The output signal is a zero crossing square wave with a fixed amplitude of 6 V peak to peak from 1 and 5 V Instantly output your SPAN and ZERO output settings by moving the EZ CHECK switch be...

Страница 31: ...aximum scale are limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit The PIE 820 ELITE measures the input signal and constantly updates the display with...

Страница 32: ...pressure instrument to be checked The following screen will appear each time pressure is selected to remind you that the switch on the Multical Pressure Module MUST be set to PSI for the PIE 820 ELIT...

Страница 33: ...the highest reading and to MIN to see the lowest reading Press and hold the knob e for 2 seconds to clear the MAX and MIN readings 30 psi 2 bar Gauge Absolute Modules Units Maximum PSI 30 000 Inches...

Страница 34: ...ight 12 1 ounces 0 34 kg with boot batteries Batteries Four AA Alkaline 1 5V LR6 Battery life Read Functions 20 hours Source mA 14 hours 12 mA into 250 Pwr Meas mA 12 hours at 20 mA SourceV T C RTD Hz...

Страница 35: ...20 00mA Loop drive capability 1200 at 20 mA for 15 hours nominal 950 with Hart Resistor enabled mA 2 Wire Transmitter Simulation Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overlo...

Страница 36: ...Compensation 0 1 C 0 01 C Output Impedance 1 Ohm Source Current 20 mA drives 80 mV into 10 Ohms Thermocouple Read Accuracy Cold Junction Compensation Same as Thermocouple Source Input Impedance 1 Mego...

Страница 37: ...mA of External Excitation Current 2 Wire Accuracy 0 03 of Full Scale 0 075 Ohms 0 03 of Full Scale 0 075 Ohms Add 0 1 Ohm to specifications listed above Resistance Ranges 0 0 to 410 00 to 4001 0 Ohms...

Страница 38: ...tude 10 microseconds Output Impedance 1 Ohm Source Current 1 mA rms at 20 kHz Short Circuit Duration Infinite Optical Coupling Green LED HZ SYNC flashes at output frequency Frequency Read Ranges Accur...

Страница 39: ...2 0 to 2500 0 1 1 T 260 0 to 230 0 2 9 436 0 to 382 0 5 2 230 0 to 210 0 1 0 382 0 to 346 0 1 9 210 0 to 50 0 0 8 346 0 to 58 0 1 4 58 0 to 50 0 0 3 58 0 to 122 0 0 6 50 0 to 400 0 0 2 122 0 to 752 0...

Страница 40: ...1 2 50 0 to 950 0 0 4 58 0 to 1742 0 0 8 950 0 to 1300 0 0 5 1742 0 to 2372 0 1 0 G W 100 0 to 350 0 1 7 212 0 to 662 0 3 0 350 0 to 1700 0 0 8 662 0 to 3092 0 1 5 1700 0 to 2000 0 1 0 3092 0 to 3632...

Страница 41: ...1395 0 0 7 2192 0 to 2543 0 1 2 L J DIN 200 0 to 50 0 0 4 328 0 to 58 0 0 7 50 0 to 300 0 0 2 58 0 to 572 0 0 4 300 0 to 900 0 0 3 572 0 to 1652 0 0 5 U T DIN 200 0 to 50 0 0 6 328 0 to 58 0 1 1 50 0...

Страница 42: ...0 2 0 3 0 4 0 5 328 0 to 68 0 68 0 to 680 0 680 0 to 1202 0 1202 0 to 1562 0 0 4 0 6 0 8 0 9 Pt 100 Ohm US Lab 1 3926 200 0 to 20 0 20 0 to 360 0 360 0 to 660 0 660 0 to 850 0 0 2 0 3 0 4 0 5 328 0 to...

Страница 43: ...ir Replacement Warranty RP WAR B our equipment is warranted against ANY damage or malfunction that may cause the unit to fail for a period of three 3 years from the date of shipment This warranty is l...

Страница 44: ...et Suite 3 14 Webster NY 14580 USA Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Copyright 2014 All rights reserved 820 E 9002 Rev B 27 May 2014 www GlobalTestSupply com Quality PI...

Отзывы: