15
[Step 4]
Insert the voltage setting selector back into the socket and close the cap as
shown in Figure 2-12.
Figure 2-12
Warning! Before re-connecting the power line cord to your multimeter,
make sure that power switch is in the “POWER OFF” state. Verify if the line
voltage setting is correct before you turn on your instrument.
2.2
Current Input Fuse – 3.15 A
There are two fuses located in the front and rear current input terminals to
protect your multimeter against damage of strong current pulse. The front
and rear current input fuses used in M3500A are the type 3.15A, 250 V, 5
×
20 mm, fast acting, HBC and Ceramic Tube with high breaking character
(Picotest part number: 024-001-000007, the reference number: FS101). If
you need to replace them, please follow the steps shown below.
Warning! For continued protection against fire or instrument damage,
only replace fuse with the same type and rating. If the instrument
repeatedly blows fuses, locate and correct the cause of the trouble before
replacing the fuse.
Содержание M3500A
Страница 1: ...PICOTEST M3500A 6 5 Digit Digital Multimeter Service Manual Printed date 11 2011 ...
Страница 52: ...52 4 5 1 Front Panel Assembly ...
Страница 53: ...53 4 5 2 Chassis and Transformer Assembly ...
Страница 54: ...54 4 5 3 Main Board Assembly ...
Страница 55: ...55 ...
Страница 56: ...56 4 5 4 Chassis Assembly ...
Страница 57: ...57 ...
Страница 82: ...82 5 3 Layout Drawings Main board PCB components layout top layer Figure 5 1 ...
Страница 83: ...83 Main board PCB components layout bottom layer Figure 5 2 ...
Страница 84: ...84 Panel PCB components layout top layer Figure 5 3 Panel PCB components layout bottom layer Figure 5 4 ...