smaxcur
<channel> <current>
--
Sets the maximum current of the selected channel
to the given value.
Note: In single channel mode the first parameter
<channel> must not be set!
gmaxcurmin
--
maximum
current in A
Returns the minimum output current of the TEC
gmaxcurmax
--
minimum
current in A
Returns the maximum output current of the TEC
ginput
<channel>
selected input
Returns the current input number of the given
channel.
0: NTC 1
1: NTC 2
3: PTC 1
4: PTC 2
Note: In single channel mode the first parameter
<channel> must not be set!
sinput
<channel> <input>
--
Sets the input of the selected channel to the given
value:
0: NTC 1
1: NTC 2
3: PTC 1
4: PTC 2
Note: In single channel mode the first parameter
<channel> must not be set!
gregler
<channel>
see text
This command prints out the internal controller
values. The format is:
<setpoint> tab <actual value> tab <err> tab <err
sum> tab <previous err> tab <duty cycle in A>
IIIIffff an Error Occurs
an Error Occurs
an Error Occurs
an Error Occurs
If an error occurs during operation the pulse output is switched off and the return value of a
command is no longer “0” or “1” but “10” or “11”. Errors have to be acknowledged with a toggle
of the ENABLE signal, otherwise switching on again of TEC output is not possible. For more details see
the description of the ERROR register.
To retrieve the error, use the gerror
gerror
gerror
gerror command for the content of the ERROR register or the gerrtxt
gerrtxt
gerrtxt
gerrtxt
command for a human readable form.