PicoScope 3000A Series Prog rammer's Guide
9
Copyright © 2011-2013 Pico Technology Limited. All rights reserved.
ps3000abpg.en r9
2.7.1.1
Using block mode
This is the general procedure for reading and displaying data in
single
1.
Open the oscilloscope using
2.
Select channel ranges and AC/DC coupling using
All channels
are enabled by default, so if you wish to allocate the buffer memory to fewer
channels, you must disable those that are not required.
3.
[MSOs only]
Set the digital port using
.
4.
Using
, select timebases until the required nanoseconds per
sample is located.
5.
Use the trigger setup functions
ps3000aSetTriggerChannelConditionsV2
ps3000aSetTriggerChannelDirections
ps3000aSetTriggerChannelProperties
to
set up the trigger if required.
6.
[MSOs only]
Use the trigger setup functions
ps3000aSetTriggerDigitalPortProperties
to set up the digital trigger if required.
7.
Start the oscilloscope running using
8.
Wait until the oscilloscope is ready using the
using
).
9.
Use
to tell the driver where your memory buffer is.
10.
Transfer the block of data from the oscilloscope using
11.
Display the data.
12.
Stop the oscilloscope using
.
13.
Repeat steps 7 to 11.
14.
Request new views of stored data using different downsampling parameters: see
2.7.1.2
Asynchronous calls in block mode
The
function may take a long time to complete if a large amount of
data is being collected. For example, it can take 3.5 seconds to retrieve the full 128M
samples from a PicoScope 3206B using a USB 2.0 connection, or several minutes on
USB 1.1. To avoid hanging the calling thread, it is possible to call
instead. This immediately returns control to the calling thread,
which then has the option of waiting for the data or calling
operation.