background image

P2036 300 MHz / P2056 500 MHz 

10:1 oscilloscope probes

User's Guide

Содержание P2036

Страница 1: ...P2036 300 MHz P2056 500 MHz 10 1 oscilloscope probes User s Guide...

Страница 2: ...re required to protect and keep our environment clean Therefore either return this product at the end of life to the manufacturer or ensure WEEE compliant collection and treatment yourself Do not disp...

Страница 3: ...should use this product only as instructed and with only accessories supplied or recommended Protection provided by the product may be impaired if used in a manner not specified by the manufacturer Ma...

Страница 4: ...A041 listed on the Pico website must be used instead WARNING To prevent injury or death do not use this probe for measurements on mains installations in measurement categories III and IV WARNING To pr...

Страница 5: ...y caused by condensation must be accepted Care of the product The probe contains no user serviceable parts Repair servicing and calibration require specialized test equipment and must only be performe...

Страница 6: ...ter visibility on the DUT device under test in dense SMT circuits than conventional 5 mm housings The remarkably low input capacitance of only 9 5 pF applies a minimal load to the signal source which...

Страница 7: ...V at DC Probe bandwidth 300 MHz 500 MHz 3 dB Probe risetime 1 ns 700 ps 10 90 typical Connected to oscilloscope with an input impedance of 1 M 1 Electrical characteristics Input resistance system 10...

Страница 8: ...ncies A poorly compensated probe influences the overall system performance probe scope and introduces measurement errors resulting in inaccurate readings and distorted waveforms LF compensation is per...

Страница 9: ...mm pitch brown IC cap 2 5 mm 1 27 mm pitch black Insulating cap 2 5 mm Solid tips CuBe 0 5 mm Spring tips gold plated 0 5 mm 2 footed probe holder Probe DEVICE UNDER TEST Set coding rings 3x4 colors G...

Страница 10: ...1 1 Ground lead 15 cm 2 2 1 1 1 1 1 Ground spring 2 5 mm 2 2 1 1 1 IC cap 2 5 0 5 mm pitch green 1 1 1 IC cap 2 5 0 65 mm pitch blue 1 1 1 IC cap 2 5 0 8 mm pitch grey 1 1 1 IC cap 2 5 1 0 mm pitch b...

Страница 11: ...s because the tip could be squeezed and made impossible to remove and the probe could be damaged After the probe tip is removed the new tip can be inserted with pliers into the contact socket along th...

Страница 12: ...13 Feb 2020 United Kingdom headquarters Pico Technology James House Marlborough Road Colmworth Business Park Eaton Socon St Neots PE19 8YP United Kingdom DO337 1...

Отзывы: