![PI Q-521 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/pi/q-521/q-521_user-manual_1554423056.webp)
10 Technical Data
52
Version: 1.0.1
MP145E
Q-521 Miniature Linear Stage
10.1.2
Maximum Ratings
The Q-521 linear stage is designed for the following operating data:
Maximum Operating
Voltage
Maximum Operating
Frequency
Maximum Power
Consumption
48 V
20 kHz
10 W
10.1.3
Ambient Conditions and Classifications
Pay attention to the following ambient conditions and classifications for the Q-521:
Area of application
For indoor use only
Maximum altitude
2000 m
Air pressure
Q-521.xx0:
1100 hPa to 10
-6
hPa
Q-521.xxU:
1100 hPa to 10
-9
hPa
Relative humidity
Highest relative humidity 80 % for temperatures up
to 31 °C, non-condensing
Decreasing linearly to 50 % relative humidity at
40 °C, non-condensing
Operating temperature
0 °C to 40 °C
Storage temperature
-20 °C to 70 °C
Transport temperature
-20 °C to 70 °C
Maximum bakeout
temperature for Q-521.xxU:
120 °C, for 12 hours, only in switched-off state
Maximum bakeout
temperature for Q-521.xx0:
80 °C, for 2 hours, only in switched-off state
Overvoltage category (acc. to
EN 60664-1 / VDE 0110-1)
II
Protection class
(acc. to EN 61140 /
VDE 0140-1)
I
Degree of pollution
(acc. to EN 60664-1 /
VDE 0110-1)
1
Degree of protection
(acc. to IEC 60529)
IP20