![PI P-753.11C Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/pi/p-753-11c/p-753-11c_user-manual_1554420031.webp)
6 Startup and Operation
P753 Positioning Systems
PZ254E
Version: 1.0.1
27
6.3
Discharging the P753
The P-753 must be discharged in the following cases:
Before installation
If the P-753 is not used and the controller remains switched on to ensure temperature
stability
Before demounting (e.g., before cleaning and transporting the P-753 and for
modifications)
The P-753 is discharged via the internal discharge resistor of the controller from PI.
Discharging a positioner that is connected to the controller
In closed-loop operation:
1.
Switch off the servo mode on the controller.
2.
Set the piezo voltage to 0 V on the controller.
In open-loop operation:
Set the piezo voltage to 0 V on the controller.
Discharging a positioner that is not connected to the controller
Connect the positioner to the switched-off controller from PI.