background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

12652-00 / 3611

 

impossible  to  guarantee  that a  product  is  free  of  faults. 
PHYWE  Systeme  GmbH  &  Co.  KG  therefore  does  not 
take  on  any  liability  for  damages  that  occur  during  the 
installation or the use of the instrument. 

2  PURPOSE AND CHARACTERISTICS 

Cobra4-Unit  Tesla  serves  to  measure  the  magnetic  field 
strengths  of  constant  and  alternating  fields  in  combination 
with a tangential or axial Hall probe. 
 
According  to  the  application,  Cobra4  Sensor-Unit  Tesla  can 
be  connected  to  one  of  the  following  measured  data  re-
cording instruments and pass measured data to it: 
 

 

to  Cobra4  Wireless-Link  (12601-00)  for  wireless  trans-

mission to a computer, 

 

to Cobra4 USB-Link (12610-00) for data transmission to 

a computer via a USB-cable, 

 

to  the  hand-held  measuring  instrument  Cobra4  Mobile-

Link (12620-00). 

 
The  interface  to  measured  data  recording  instruments  is  on 
the  front  side  of  the  Sensor-unit,  whereby  mechanically  se-
cure  engagement  is  ensured  by  a  locking  mushroom  and  a 
hole.  
 
For  measurements  of  magnetic  field  strengths,  a  tangential 
(13610-02)  or  an  axial  (13610-01)  Hall  probe  can  be  con-
nected to the Sensor-Unit, according to the application. It has 
a 5-pin DIN socket for connection of the Hall probe.  

3  FUNCTIONAL AND OPERATING ELEMENTS 

3.1  Operating elements

 

Cobra4  Sensor-Unit  Tesla  has  no  manual  operating  ele-
ments.  Operation  of  it  is  made  via  the  Cobra4  Mobile-Link 
hand-held measuring instrument, or via a computer to which 
either  a  Cobra4  USB-Link  or  a  Cobra4  Wireless-Link  has 
been connected, as soon as it has been connected to the se-
lected  one  of  the  three  allowable  measured  data  recording 
instruments. 

4  NOTES ON OPERATION  

Together  with  Cobra4–Wireless-Link,  Cobra4–USB-Link  and 
Cobra4–Mobile-Link,  Cobra4–Sensor-Unit  Tesla  fulfils  the 
technical requirements that are compiled in current European 
Community  Guidelines.  The  characteristics  of  this  product 
qualify it for the CE mark. 
This instrument is only to be operated under specialist super-
vision  in  a  controlled  electromagnetic  environment  in  re-
search,  educational  and  training  facilities  (schools,  universi-
ties, institutes and laboratories). 
This  means  that  in  such  an  environment  no  mobile  phones 
etc.  are  to  be  used  in  the  immediate  vicinity.  The  individual 
connecting leads are each not to be longer than 2 m. 
 
The instrument can be so influenced by electrostatic charges 
and other electromagnetic phenomena that it no longer func-
tions  within  the  given  technical  specifications.  The  following 
measures reduce or do away with disturbances: 
Avoid  fitted  carpets;  ensure  potential  equalization;  carry  out 
experiments on a conductive, earthed surface, use screened 
cables,  do  not  operate  high-frequency  emitters  (radios,  mo-
bile phones) in the immediate vicinity.  

In  combinations  of  the  Sensor-Unit  with  USB-Link 
(12610.00), EM-disturbances can end the measurement pro-
gramme.  A  restart  (reset)  of  the  device  must  then  made  by 
unplugging  and  re-plugging  the  USB-cable.  Measurement 
must then be started anew in the “Measure” programme. 
In extreme cases, USB-communication can be re-established 
by restarting the computer. 

5  HANDLING 

This section describes how to put the Sensor-Unit into opera-
tion and record measured data. Please read carefully through 
it to avoid failure or faulty operation. 

 
5.1  Putting the Cobra4 Sensor-Unit Tesla into operation 

Check  that  the  current  version  of  the  measure  software 
package and the Cobra4 module are installed. 
Updating of the software is free via: 
www.phywe.com… software … measure 
 

measure main programme 

 

measure Cobra4 module 

Free  updating  of  the  software  can  also  be  made  with  the 
measure DVD. 
 
Connect  your  Cobra4  Sensor-Unit  Tesla  to  one  of  the  three 
measured data recording instruments via the engaging 15-pin 
plug connection; the Sensor-Unit is ready to use after a brief 
starting phase (self-calibration), as voltage is supplied by the 
measured data recording instrument. 
Further  information  on  the  control  of  the  measured  data  re-
cording instruments is given in their operating instructions..

 

6  TECHNICAL SPECIFICATIONS 

(Typical for 25°C) 

Operating temperature range 5...40°C 
Relative humidity < 80% 
 
Measuring range, constant field 
 

-

 

± 1 

; Resolution ± 1 mT 

-

 

± 100 mT  ; Resolution ± 100 µT 

-

 

±   10 mT  ; Resolution ±   10 µT   

-

 

Each measurement range can be 
compensated up to the final value of 
the particular measurement range. 

-

 

Accuracy approx. ± 2 % of measure-
ment range final value 

 

 
Measuring range, alternating field from 15 Hz to 1 kHz  
 

-

 

1 T 

 

; Resolution  1 mT 

-

 

100 mT 

; Resolution  

100 µT 

-

 

10 mT 

; Resolution 

  10 µT 

-

 

Alternating fields are measured with a 
root-mean-square rectifier (RMS) 

-

 

No measurement range compensation 

-

 

Accuracy approx. ± 5 % of measure-
ment range final value 
 

Scan rate, maximum (Hz) 

Dimensions (L x W x H in mm) 

60 x 70 x 35 

Weight (g) 

100 