3  FUNCTIONAL AND OPERATING ELEMENTS 

3.1 Operating elements 

The sensor has an on-button and two LEDs for indicating the 
Bluetooth and battery charge status. 
 

On-button 

To switch the sensor on and off in Bluetooth mode, the power 
button must be pressed for longer than 3s. If the sensor is to 
be connected via USB, it is not necessary to press the power 
button. 

Bluetooth-LED  

Flashing red every 2 seconds 

Not connected 

Flashing green every 2 seconds 

Connected to the ter-
minal device 

Flashing green every 4 seconds 

Running measurement 

 

Battery charge LED 

Flashing red every 2 seconds 

Low battery 

Illuminated red 

Active charging process 

Illuminated red 

Charging process 
completed 

 
3.2 USB port 

The  battery,  which  is  permanently  installed  in  the  sensor,  is 
charged  via  the  type  C USB  port.

 

Furthermore,  communica-

tion with a computer takes place via this interface. 

4  NOTES ON OPERATION 

For safety reasons, the Cobra SMARTsense sensor may only 
be used when the sensor is not connected to mains voltage! 
To do this, make sure that either 

a)  The Cobra SMARTsense sensor is not connected to 

a charger via USB. 

b)  If  a  USB  connection  to  a  laptop  is  established,  it  is 

used in battery mode with the power supply discon-
nected, or 

c)  If  there  is  a  USB  connection  to  a  standard  PC  or 

laptop  with  a  power  supply  unit,  these  are  con-
nected  to  the  mains  voltage  via  a  floating  isolating 
transformer. 

 
This    device  fulfils  all  of  the  technical  requirements  that  are 
compiled in current EC guidelines. The characteristics of this 
product qualify it for the CE mark. 
This instrument is only to be put into operation under special-
ist supervision in a controlled electromagnetic environment in 
research, educational and training facilities (schools, universi-
ties, institutes and laboratories). 
 The  individual  connecting  leads  are  each  not  to  be  longer 
than 2 m. 
The instrument can be so influenced by electrostatic charges 
and  other  electromagnetic  phenomena  (HF,  bursts,  indirect 
lightning  discharges)  that it  no  longer  works  within  the  given 
specifications. Carry out the following measures to reduce or 
eliminate  the  effect  of  such  disturbance:  Ensure  potential 
equalization at the PC (especially with Laptops). Use screen-
ing.  When  a  total  failure  of  the  instrument  occurs,  unplug  it 
and plug it back in again for a reset. 
 

5  HANDLING 

This section describes the start-up of the sensor and the re-
cording of measurement  data.  Please  read this section  thor-
oughly in order to avoid failures or operating errors. 

 
5.1  Charging process 

Use  a  USB-C  cable  to  connect  the  sensor  to  a  computer  or 
USB charger (not included). 
During  the  charging  process,  the  battery  charge  LED  lights 
up  red.  When  the  charging  process  is  complete,  the  battery 
charge  LED  lights  up  green.  The  charging  time  for  a  com-
pletely discharged battery is 3 hours maximum. 
 

Disconnect the charger at the latest four hours af-
ter the completion of the charging process. Other-
wise,  the  service  life  of  the  battery  may  be  nega-
tively affected. 

 
5.2  Start-up 

Switch the sensor on by pressing the on-button for more than 
3  seconds.  The  Bluetooth  LED  lights  up  red.  Start  the  soft-
ware and select the sensor. 
 

9-digit 

code 

is 

printed  on  the  back  of 
the sensor (Fig.2). The 
last 4 digits of the code 
are 

displayed 

as 

sensor  designation  in 
the  software  (Fig.3). 
This  enables  an  exact 
assignment 

of 

the 

sensors 

with 

the 

software.  
 

   

 

 

 

 

Fig. 2 

 
 
 

   

 

 

 

   

 

 

   

 

 

Fig. 3 
 

Selection of the sensor via the Bluetooth interface 

Make  sure  that  the  Bluetooth  interface  is  activated  on  the 
terminal  device  (PC/Tablet/Smartphone)  and  that  the  soft-
ware is allowed to access the interface.  
 
After the sensor has been selected in the software, the LED 
flashes green to indicate that the connection has been estab-
lished  correctly.  After  the  sensor  has  been  coupled  with  the 
software, the sensor is no longer visible to other users in the 
software, and therefore can no longer be selected.  
 
If the sensor is switched on and not connected, it switches off 
automatically after 5 minutes. 
 

Selection of the sensor via the USB interface 

For  this  purpose  the  sensor  must  be  plugged  into  the  USB 
port  of  the  end  device.  It  is  not  necessary  to  switch  on  the 
sensor.  The  sensor  is  automatically  recognized  and  dis-
played. It can be selected and connected directly. 
 
 
 
 
 
 
 
 