3  FUNCTIONAL AND OPERATING ELEMENTS 

3.1 Operating elements 

The sensor has an on-button and two LEDs for indicating the 
Bluetooth and battery charge status. 
 

On-button 

Press  the  on-button  for  more  than  3  seconds  to  switch  the 
sensor on and off 
 

Bluetooth-LED  

Flashing red every 2 seconds 

Not connected 

Flashing green every 2 seconds 

Connected to the ter-
minal device 

Flashing green every 4 seconds 

Running measurement 

 

Battery charge LED 

Flashing red every 5 seconds 

Low battery 

 
3.2 Measurement inputs

 

The  front  of  the sensor  is  equipped  with  a fixed  conductivity 
measuring cell with an integrated temperature sensor. 

4  NOTES ON OPERATION 

This  device  fulfils  all  of  the  technical  requirements  that  are 
compiled in current EC guidelines. The characteristics of this 
product qualify it for the CE mark. 
 The  individual  connecting  leads  are  each  not  to  be  longer 
than 2 m. 
The instrument can be so influenced by electrostatic charges 
and  other  electromagnetic  phenomena  (HF,  bursts,  indirect 
lightning  discharges) that  it  no  longer  works  within  the  given 
specifications. Carry out the following measures to reduce or 
eliminate  the  effect  of  such  disturbance:  Ensure  potential 
equalization at the PC (especially with Laptops). Use screen-
ing.  Do  not  operate  high  frequency  emitters  (e.g.  radio 
equipment or mobile radiotelephones) in the immediate vicini-
ty. When a total failure of the instrument occurs, unplug it and 
plug it back in again for a reset. 

5   HANDLING 

This section describes the start-up of the sensor and the re-
cording  of  measurement  data.  Please  read this  section thor-
oughly in order to avoid failures or operating errors. 
 

5.1  Start-up 

Switch the sensor on by pressing the on-button for more than 
3  seconds.  The  Bluetooth  LED  lights  up  red.  Start  the  soft-
ware and select the sensor. 
 
There  is  a  9-digit  code 
on the back of the sen-
sor  (Fig.2).  The  last  4 
digits  of  the  code  are 
displayed as the sensor 
name  in  the  software 
(Fig.3).  This  enables 
the  precise  assignment 
of  the  sensors  within 
the software. 
 

Fig. 2 

 
 

 
 
 
 
 

Fig. 3 

 
5.2 Recording of conductivity values  

To measure the conductivity of the various media, the meas-
uring  head  of  the  conductivity  measuring  cell  is  completely 
plunged into the medium to be measured. If a whole series of 
different solutions is to be measured, we recommend starting 
with the  solution with the  lowest  conductivity.  Since  some  of 
the  sample  material  is  always  carried  over  to  the  next  sam-
ple, this method ensures that a potential error remains on an 
acceptable  level  as  the  contamination  of  a  solution  with  a 
high  level  of  conductivity  with  traces  of  a  solution with  a  low 
level of conductivity leads to a smaller relative error than the 
other way round. In any case, the measuring head and elec-
trodes  must  be  rinsed  with  pure  (demineralised  or  distilled) 
water when switching from one sample to another. 
The  solution  should  always  remain  slightly  in  motion  during 
the  measurement.  This  can  be  realised  by  stirring  the  liquid 
with a magnetic stirrer or similar or by moving the measuring 
probe carefully to and fro. 

 

5.3 Maintenance and cleaning of the measuring probe  

In  general,  conductivity  measuring  cells  are  not  subject  to 
ageing.  However,  the  measuring  electrode  can  be  corroded 
and,  thereby,  damaged  by  certain  media  (e.g.  strong  acids 
and  lyes).  This  is  why  all  of  the  measuring  cells  should  be 
rinsed thoroughly with clean (demineralised or distilled) water 
whenever they have been used. If the measuring electrode is 
contaminated with grease or oil, it can be cleaned with warm 
water and a commercially available household detergent. 
The  measuring  cell  comes  supplied  with  a  protective  cap 
which is filled with distilled water. Do not allow the measuring 
cell to dry out. This might cause the platinum black to crum-
ble off which, in turn, would change the cell constant. This is 
why it should always be stored in pure water. When cleaning 
the  measuring  cell,  avoid  any  mechanical  stress  on  the 
measuring electrodes. 
 

 

 

5.4 Calibration 

In  general,  the  conductivity  measuring  cell  does  not  need  to 
be calibrated if it is maintained properly. 
If  a  calibration  needs  to  be carried  out,  it  is  a two-point  cali-
bration.  The  calibration  data  are  entered  via  the  measure-
ment software (App). 
 
First calibration point 
Dip  the  measuring  cell  into  distilled/deionised water  or  ultra-
pure water. 
 

 

Conductivity at 25°C 

Ultra-pure water 

0.055 µS/cm 

Deionised water 

1 µS/cm 

 
 
 
Enter  the  value  (see the  table)  as set  point  1  and  send  it  to 
the sensor. 

Measuring electrodes 

Protective cap 