and almost instant response of the probe. Fluctuations can easily be eliminated by
bringing the probe into contact with a metallic object, thus increasing its effective
mass and slowing the response. Using a larger probe will have the same results.

6.10 

Measurements on Solid Surfaces

. These are most easily made with surface

probes such as our BT-1 and MT-D. The right-
angled tip provides the 10 diameters of probe
contact specified in Section 6.6. Straight probes
may also be used, provided that sufficient shaft
length is in contact with the surface to be meas-
ured. In general, the smaller the probe, the more

accurately it will measure the surface temperature of a solid.  For instance, an MT-
29 microprobe, because of its small size, needs to be in contact with the surface for
as little as 1/8”.  SST-1 has a 1/4” gold disc sensor.  Gold is an excellent conductor,
and is non-allergenic and non-polluting.  It makes a fine skin surface probe.

6.11

Measurements in Electronics. 

BAT-12 and ICT-4 provide the quickest and

cheapest method of getting accurate reading on both discreet components and inte-
grated circuits used in electronic equipment. ICT-4 has two important differences
from the regular MT-4 probe. There is a 5 1/2” handle suitable for use in the hand or
in a micromanipulator and the micro-thermocouple protrudes slightly beyond the
steel sheath and has some thermal isolation from it. This probe was designed for use
on areas of 5-100 microns in diameter and is used perpendicular to the surface to be
measured. The heat loss described in 6.3 occurs and temperature readings are slight-
ly low  --  normally about 0.5°C low. Each probe can be readily calibrated on a larg-
er surface. First lay it parallel to the surface and note the reading. Then take a read-
ing with it perpendicular to the surface; note the difference. Add this difference to
the subsequent reading when using the perpendicular position.

6.12 BAT-12 with a suitable probe has the ability to measure surface gradients,
giving almost instant readings at each point. This feature enables heat sinks to be
designed without guesswork, for maximum efficiency with minimum metal. It is
equally easy to check the adequacy of thermal bonding between a power transistor
or rectifier and its heat sink. General purpose probe BT-1 is suitable for this and
many other measurements. When probes such as BT-1 and the larger HT-1 and HT-2
are used for surface measurements, speed of response and accuracy can often be
improved by use of a heat conductive material such as ‘thermal grease’ or silicone
grease applied between the probe and surface. Note: BT-1, HT-1 and HT-2 should

8

Содержание BAT-12

Страница 1: ...ODUCTION 2 0 INITIAL INSTALLATION 3 0 OPERATING INSTRUCTIONS 4 0 BATTERY REPLACEMENT 5 0 ANALOG OUTPUT 6 0 TEMPERATURE MEASUREMENT WITH THERMOCOUPLE SENSORS 7 0 TROUBLE SHOOTING 8 0 CHECKING CALIBRATI...

Страница 2: ......

Страница 3: ...the same box if small or a separate package if they are longer than 12 inches 2 2 BAT 12 for battery operation is supplied ready for use with a 9V alkaline bat tery already installed Shipment will in...

Страница 4: ...read temperature 4 0 BATTERY REPLACEMENT 4 1 When a colon appears in the center of the ditigal display the battery needs to be replaced However BAT 12 can be used without significant calibration error...

Страница 5: ...he analog output with strip chart recorder The output impedance of BAT 12 is 2 Kilohms so connection to a potentio metric recorder will not affect either the calibration or accuracy The analog output...

Страница 6: ...ronic reference circuitry 6 3 As compared with thermistor sensors which were formerly used exclusively in portable thermometers thermocouples have these advantages a Wide temperature range e g 200 C t...

Страница 7: ...nical probes are guaranteed accurate to within 0 1oC in the range 0 50oC Copper Constantan type T thermocouples from other manufacturers are not normally this accurate Probes made from wire to special...

Страница 8: ...1 2 handle suitable for use in the hand or in a micromanipulator and the micro thermocouple protrudes slightly beyond the steel sheath and has some thermal isolation from it This probe was designed f...

Страница 9: ...og output Remove the output plug d Probe or sensor may be faulty open circuit Check by substituting anoth er probe or by a resistance check All probes should read less than 2000 ohms If you do not hav...

Страница 10: ...ter can be invalidated by overheating ensure that this never happens 8 3 Bind the thermocouple probe to the tip of the glass thermometer so that they are close together Do not compress the glass therm...

Страница 11: ...d Technology tables NIST monograph 125 Ambient Cold Junction 0 1 C from 0 to 50 C Temp Compensation Readout 3 1 2 digits 1 2 liquid crystal numerals Analog output 10mV per degree C approximately Power...

Страница 12: ...itemp aol com 10 2 Physitemp Instruments Inc warrants this system to be free from defects in material or workmanship for 12 months from date of shipment Repair or replace ment will be made at no charg...

Страница 13: ...PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton NJ 07013 U S A Tel 973 779 5577 Fax 973 779 5954 e mail physitemp aol com OPERATING MANUAL BAT 12 Microprobe Thermometer...

Отзывы: