OPERATING INSTRUCTIONS FOR BAT-12 THERMOMETER

1.0 INTRODUCTION

The Physitemp BAT-12 thermometer is a compact, portable, digital thermometer for
use in the lab or in the field. It can be used with any Type T thermocouple sensor and
is available in two versions:

BAT-12

, for battery operation only, uses a 9V alkaline battery.

BAT-12R

, a rechargeable version, is supplied with a 9V Nickel/Metal-Hydride

battery and may be battery operated in the field or AC operated in the lab. A 
heavy duty vinyl carrying case is supplied with both versions.

2.0 INITIAL INSTALLATION

2.1

Remove the packing list (usually found in a transparent envelope on the out-

side of the carton) and unpack the instrument. Check that all items shown on the
packing list are present. Sensor probes ordered with the instrument will be packed in
the same box if small or a separate package if they are longer than 12 inches.

2.2

BAT-12, for battery operation, is supplied ready for use with a 9V alkaline bat-

tery already installed. Shipment will include the carrying case and a male plug for
connection to the analog output.

2.3

BAT-12R for AC and battery operation, is supplied with a 9V Nickel/Metal-

Hydride battery installed. This is shipped with a full charge. To recharge or for AC
use, connect the small phono plug on the AC adaptor to the jack on the rear of the
instrument and plug the charger into a live 115V outlet. 14 hours are required for a
complete recharge. Overcharging will not harm the unit. If requested, an adaptor for
220VAC operation can be supplied. A carrying case and male plug for connection to
the analog output are supplied with
the BAT-12R.

2.4

If Tilt Stand TTS-2 has been

ordered, mount the instrument on the
stand by attaching the ball and socket
mount to the threaded fitting on the
underside of the case.

3

TTS-2 TILT STAND

Содержание BAT-12

Страница 1: ...ODUCTION 2 0 INITIAL INSTALLATION 3 0 OPERATING INSTRUCTIONS 4 0 BATTERY REPLACEMENT 5 0 ANALOG OUTPUT 6 0 TEMPERATURE MEASUREMENT WITH THERMOCOUPLE SENSORS 7 0 TROUBLE SHOOTING 8 0 CHECKING CALIBRATI...

Страница 2: ......

Страница 3: ...the same box if small or a separate package if they are longer than 12 inches 2 2 BAT 12 for battery operation is supplied ready for use with a 9V alkaline bat tery already installed Shipment will in...

Страница 4: ...read temperature 4 0 BATTERY REPLACEMENT 4 1 When a colon appears in the center of the ditigal display the battery needs to be replaced However BAT 12 can be used without significant calibration error...

Страница 5: ...he analog output with strip chart recorder The output impedance of BAT 12 is 2 Kilohms so connection to a potentio metric recorder will not affect either the calibration or accuracy The analog output...

Страница 6: ...ronic reference circuitry 6 3 As compared with thermistor sensors which were formerly used exclusively in portable thermometers thermocouples have these advantages a Wide temperature range e g 200 C t...

Страница 7: ...nical probes are guaranteed accurate to within 0 1oC in the range 0 50oC Copper Constantan type T thermocouples from other manufacturers are not normally this accurate Probes made from wire to special...

Страница 8: ...1 2 handle suitable for use in the hand or in a micromanipulator and the micro thermocouple protrudes slightly beyond the steel sheath and has some thermal isolation from it This probe was designed f...

Страница 9: ...og output Remove the output plug d Probe or sensor may be faulty open circuit Check by substituting anoth er probe or by a resistance check All probes should read less than 2000 ohms If you do not hav...

Страница 10: ...ter can be invalidated by overheating ensure that this never happens 8 3 Bind the thermocouple probe to the tip of the glass thermometer so that they are close together Do not compress the glass therm...

Страница 11: ...d Technology tables NIST monograph 125 Ambient Cold Junction 0 1 C from 0 to 50 C Temp Compensation Readout 3 1 2 digits 1 2 liquid crystal numerals Analog output 10mV per degree C approximately Power...

Страница 12: ...itemp aol com 10 2 Physitemp Instruments Inc warrants this system to be free from defects in material or workmanship for 12 months from date of shipment Repair or replace ment will be made at no charg...

Страница 13: ...PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton NJ 07013 U S A Tel 973 779 5577 Fax 973 779 5954 e mail physitemp aol com OPERATING MANUAL BAT 12 Microprobe Thermometer...

Отзывы: