Title
LPCR2 Service Process Work Instruction
Page
4 of 23
Doc Type
Service Process Document
Doc ID
3323299
Doc Rev
C
Doc State
Released
State Date
2/26/2020
Ref. Product DMR
:
Ref. PFMEA:
Ref. PFMEA Rev:
Template: 7000743_B
PCO file
Physio Case Object file, a data format for a combination of device data and user-entered
data. Data in this format can be imported and exported by Physio-Control data management
applications.
SmartDesk
Data Solutions Technical Support
WCT
Wireless Configuration Tool is a software Application to configure Wi-Fi connection in the
LIFEPAK CR2® Defibrillator.
Workplace and
Community
A Physio-control customer segment that represents public access locations for Automated
External Defibrillators such as schools, government agencies, airports, health clubs, and
workplaces.
Tools Required: (substitution of equivalent or better equipment allowed)
Computer
– Win 7, equivalent or better.
Cable, USB2.0 A Male to Micro-B, L 5.5 ft. (21300-008143)
Torque Driver
– Hand Operation, with an adjustable range of 5 – 12 in-lbs.
Driver Bits
– P1 (2”minimum shaft), P2 (5” minimum shaft), P0 power drive bit,Torx, T-15 (2” minimum shaft).
Tools defined in the PIP - TCP sections of Test Procedure, Manual PIP, LPCR2, 3323342.
•
Torque Driver conversions
–
o
2 in-lbs
. =22.5 cNm
o
4 in-lbs.
= 45 cNm
o
6.8 in-lbs.
= 77 cNm
o
10 in-lbs.
= 113 cNm